Chapter 3 — pH Calibration Extension
29
The
buttons and
pop-up menu next to the axes can also be
used for scaling.
Manual Compensation Response
When the manual temperature compensation mode is selected and a
temperature other than the calibration temperature is entered, the
electrode response dialog will display two electrode plots: one
representing the response at the calibration temperature; and a second
plot showing the response at the specified temperature.
The calibration response can be identified as the line with one or two
calibration points on it. These are represented as dots. The information
displayed on the right of the dialog shows the calibration points and
the values used for the Nernst equation.
Calibration point
Calibration
points
Single point calibration uses
an ideal Nernstian slope
Double point calibration
Calibration points
Print
calibration
report
Copy
calibration
report
Figure 3–11
The Electrode Response
dialog box