SE-1010 PC ECG User Manual Operation Instructions for Resting ECG
- 52 -
When the display mode is set to 3*1, 3-channel ECG waves are displayed on one screen
simultaneously.
When the display mode is set to 3*4+1, 12-channel ECG waves are displayed in 4 groups of 3
and one rhythm lead on one screen.
When the display mode is set to 3*4+3, 12-channel ECG waves are displayed in 4 groups of 3
and three rhythm leads on one screen.
When the display mode is set to 6*2+1, 12-channel ECG waves are displayed in 2 groups of 6
and one rhythm lead on one screen.
6.5.2 Specifying Lowpass Filter
Lowpass Filter restricts the bandwidth of input signals. The cutoff frequency can be set to 25Hz,
35Hz, 45Hz, 75Hz, 100Hz, or 150Hz. The input signals whose frequency is higher than the set
cutoff frequency will be attenuated.
6.5.3 Specifying Gain
You can set the indicated length of 1mV ECG on the paper.
You can set the gain to 2.5mm/mV, 5mm/mV, 10mm/mV or 20mm/mV.
6.5.4 Specifying Speed
You can set the paper speed to 5mm/s, 10mm/s, 12.5mm/s, 25mm/s or 50mm/s.