SE-301 Series Electrocardiograph User Manual Operation Preparations
- 16 -
Chest Electrode Connection:
1)
Ensure that the electrodes are clean;
2)
Align all lead wires of the patient cable to avoid twisting, and connect the lead wires to
the corresponding electrodes according to the colors and identifiers;
3)
Clean the electrode area on the chest surface with 75% alcohol;
4)
Daub the round area of 25mm in diameter on each electrode site with gel evenly;
5)
Place a small amount of gel on the brim of chest electrode’s metal cup;
6)
Place the electrode on the chest electrode site and squeeze the suction bulb. Unclench it
and the electrode is adsorbed on the chest;
7)
Attach all chest electrodes in the same way.
NOTE:
Long-time measurement with a strong negative pressure on the suction bulb
may cause reddening of the skin. When using the electrode on small children or
patients with delicate skin, squeeze the suction ball lightly.
Limb Electrode Connection:
2)
Ensure that the electrodes are clean;
3)
Align all lead wires of the patient cable to avoid twisting,
and connect the lead wires to the corresponding
electrodes according to the colors and identifiers;
4)
Clean the electrode area which is a short distance above
the ankle or the wrist with alcohol;
5)
Daub the electrode area on the limb with gel evenly;
6)
Place a small amount of gel on the metal part of the limb
C6
C1
C2
C5
C4
C3