H100B Pulse Oximeter User Manual
- 47 -
PR Alarm is from “1 + the lower limit of PR
Alarm” to 300.
Press the
Up Arrow
or
Down Arrow
button to
increase or decrease values.
Choose
PR Lo Alarm
in menu, press the
Function
button enter setup. The
PR Lo
Alarm
box changes from real line to broken
line. The adjustable range for the lower limit
of PR Alarm is from 0 to “the upper limit of
PR Alarm – 1”.
Press the
Function
button, confirm the alarm
range setup.
Hi Alarm
is always higher than
Lo Alarm
by
at least 1 bpm.
Press the
On/Off
button to exit the menu, and
return to measurement state.
3.3.12 Patient ID No. Setup
The oximeter can support 100 patient IDs, and 300-hour
data storage.
When entering menu, press the
Function
button to set
ID (valid range is from 1 to 100). The ID display box on
the interface will change from real line to broken line.