H100B Pulse Oximeter User Manual
- 11 -
10
A functional tester cannot be used to assess
the accuracy of the pulse oximeter probe or the
pulse oximeter monitor.
11
If there is independent demonstration that the
particular calibration curve is accurate for the
combination of a pulse oximeter monitor and a
pulse oximeter probe, then a functional tester
can measure the contribution of a monitor to the
total error of a monitor/probe system. The
functional tester can then measure how
accurately a particular pulse oximeter monitor is
reproducing that calibration curve.
12
The operating time of the Ni-MH rechargeable
battery package depends on the configuration
and operation of the pulse oximeter.
1.4 Symbols in the Oximeter
This symbol indicates that the
instrument is IEC/EN 60601-1
Type BF equipment.