Section 4: Preventive Maintenance
Cleaning the Unit
For cleaning and disinfecting always use conventional hospital-approved, topical equipment cleaners, and disinfectants.
Thoroughly wipe down device with a damp cloth to remove any residue from cleaning solutions. Ensure that all
inaccessible cracks and crevices on the WarmAir are reached during cleaning.
Hour Meter
The WarmAir
135 unit is equipped with a built-in timer that will activate the "Hour Meter" light after 500 hours of use.
This is an indication that routine maintenance is required. See Operation & Technical Manual for instructions. Instructions
on replacing the WarmAir
135 air filter are also located in the Operation & Technical Manual.
WarmAir
135 unit has approval in accordance with IEC 80601-2-35. These standards are based on contact surface
temperature of the blanket being 48°C or less. Temperatures referenced at the hose outlet do not represent
contact surface temperatures of the blanket due to temperature loss through the hose and dispersion through the
blanket.