CAUTION!
High temperature setting to be used with close patient observation when treating patients
with the following conditions:
a. Significant peripheral vascular disease, occlusive or diabetic in nature.
b. Low cardiac output.
c. Do not apply to ischemic limbs, e.g., during arterial cross-clamping.
CAUTION!
Do not attempt to warm patient with hose only. Operating the WarmAir
135 unit without a
blanket may result in patient injury.
CAUTION!
High temperature setting to be used with close patient observation. Reduce air temperature
or discontinue therapy when the therapeutic goal is reached or if vital signs instability
occurs.
Over-Temperature Safety Limit. This LED indicator light will indicate an over-
temperature condition (an audible alarm will also sound). Immediately
discontinue use and remove from service if activated.
Under Temperature Safety Limit. This LED indicator light will indicate an under-
temperature condition (an audible alarm will also sound). Immeidately
discontinue use and remove from service if activated.
Hour Meter. This LED indicator will alert the caregiver that 500 hours of
service has transpired and the unit is due for its regular preventative
maintenance, including changing the filter.
!
!
!
!