Transport and handling 

 

The safety cabinet that you received has been protected and transported in optimal conditions, as to be protected from any damage.  

It has to be lifted only by suitable equipment (suct

ion cup, strap…). It has to be carried in a vertical position, and it must not be lying.  

It is imperative not to tilt the cabinet when it is carried. Indeed it can cause damages that can alter the right functioning of the cabinet.  

On the other hand, if it has to be moved, the cabinet has to be placed on a pallet and strapped, as to avoid any fall.  

 

Commissioning 

 

After having unwrapped the cabinet, ensure that you also received a set of two keys.  

Your safety cabinet has to be installed, used and maintained as to guarantee the users protection.  

You need to remove the covers on the vents before storing any can in your cabinet. To do this, simply unscrew them counterclockwise.  

The doors open by pulling the handle towards you.  

 

4.  INTERNAL EQUIPMENTS 

 

The cabinet is provided with 2 shelves. This must be fixed at the chosen height thanks to the rack and pinion.  

Take care not to exceed the loading capacity per shelf of 159 KG.  

The cabinet also has an integrated retention tank, which has a retention capacity of 40 liters.  

 

5.  STORAGE 

 

Your safety cabinet is exclusively destined to store chemicals in laboratories, industries, schools

… in closed containers.  

It must not store high concentrated corrosive products.  

We remind  you  that  the  storage  of corrosives  can  alter  the  quality  of  your safety  equipment.  On  the  other  hand,  we  remind  you  that 

damages due to a bad use of the equipment and in particular to corrosion won

’t be covered by our warranty.  

 

6.  VENTILATION 

 

We highly recommend connecting your safety cabinet to a ventilation system, as to evacuate the eventual vapors contained in it.  

The air renewal has to be at least 10 times the volume of the cabinet, per hour. Moreover, ensure that the ventilation runs 24h/7.  

 

Connection to a ventilation system 

The air inlet and outlet are situated on the side of the cabinet. See the schema below.  

 
 

Summary of Contents for MC90

Page 1: ...Range 2 MC90 Operating instructions ECOSAFE Date du document 29 04 2019 Version 01 0 Instruction manual For safety cabinet Model MC90 ...

Page 2: ...ONTENT 1 Description Use Page 2 2 Technical features Page 2 3 Installation and commissioning Page 2 4 Internal equipments Page 3 5 Storage Page 3 6 Ventilation Page 3 7 Safety controls Page 4 8 Spare parts Page 5 9 Recycling Page 5 10 Warranty Page 5 11 Control sheet for the user Page 7 ...

Page 3: ...ters for versatile storage Leakproof 2 51 mm bottom sump Dual vents with flame arresters 4 self leveling feet Grounding connector External dimensions H 1650 x W 1100 x D 860 mm Internal dimensions H 1560 x W 1020 x D 780 mm Storage capacity 340 liters Retention capacity 40 liters Loading capacity 159 KG per shelf Weight 220 KG 3 INSTALLATION AND COMMISSIONING Installation The safety cabinet has to...

Page 4: ...se The doors open by pulling the handle towards you 4 INTERNAL EQUIPMENTS The cabinet is provided with 2 shelves This must be fixed at the chosen height thanks to the rack and pinion Take care not to exceed the loading capacity per shelf of 159 KG The cabinet also has an integrated retention tank which has a retention capacity of 40 liters 5 STORAGE Your safety cabinet is exclusively destined to s...

Page 5: ... a healthy fashion Filtration boxes are delivered with the activated carbon filter that you need to renew once a year Running without any mechanic ventilation It is possible for your safety cabinet to run without mechanical ventilation However your cabinet has to be placed in a location that provides its own ventilation system at least 5 times the volume of the location per hour The cabinet itself...

Page 6: ...arious elements such as the steel can be recycled Ensure to respect the national and local waste disposal terms Those elements should not be mixed with household waste 10 WARRANTY Your equipment holds a 12 months warranty against every manufacturing defect limited to ordinary conditions of use from the date indicated on the delivery note Our warranty can be applied only in case of manufacturing de...

Page 7: ...mal attrition of the equipment In case of a bad installation of the equipment In case of damage caused by the corrosion The activation of the warranty won t extend the term Our warranty stops if the customer has not notified us in a 7 days term after seeing the fault date that the customer has to prove ...

Page 8: ...y The locking system of the doors works correctly The mechanic ventilation system is running correctly if applicable The ventilation system is not obstructed The air inlet and outlet are clean The standardized pictogram on the door is in good condition and visible The door hinges are lubricated The locking system is lubricated The containment sump is clean Remarks Date By Signature ...

Reviews: