1.  DESCRIPTION / USE 

 

The corrosive products (acids and bases) have to be stored in safety cabinets to avoid any risk of leak or breakage of a bottle. Please 

ensure to separate acids and bases storage considering their incompatibility. 

These  safety  cabinets  are  compliant  to  the  standard  FM  and  to  the  NFPA 

–  OSHA  specifications  with  manual  or  automatic  closing 

doors. They allow storing either acids or bases, with a protection at every storage level in polyethylene. 

 

2.  TECHNICAL FEATURES 

 

Compliance :  

 

Compliance  with  the  specifications  NFPA  (National  Fire  Protection  Association),  OSHA  (Occupational  Safety  and  Head 

Administration) and FM (Factory Mutual International) 

 

Technical features :  

 

Construction in 18-gauge (1-mm) thick double-wall, welded steel with 1-

1/2” (38-mm) of insulating air space 

 

Standardized pictogram for acids provided (trilingual) 

 

Swing door 

 

Manual door, that open to a full 180 degrees 

 

3-point locking-system 

 

Shelves adjustable 

on 3” (76-mm) centers for versatile storage 

 

L

eakproof 2” (51-mm) bottom sump 

 

Dual vents with flame arresters 

 

4 self-leveling feet 

 

Grounding connector 

 

External dimensions : H 910 x W 890 x D 560 mm 

 

Internal dimensions : H 710 x W 795 x D 455 mm 

 

Storage capacity : 90 liters 

 

Retention capacity : 18 liters 

 

Loading capacity : 75 KG per shelf 

 

Weight : 92 KG 

 

3.  INSTALLATION AND COMMISSIONING 

 

Installation 

 

The safety cabinet has to be installed on a flat surface that can bear its weight with the products inside.  

Ensure that the safety cabinet is correctly wedged, as for the cabinet not to topple. The eventual blocking has to be done by the user.  

The correct leveling of the cabinet must cart off any distortion of the frame, allow the doors to be strictly parallel compared to the frame, 

and take away any rubbing of the doors on the cabinet’s frame.  

 

The installation location must guarantee :  

 

That it is not exposed to any heat source 

 

That it is protected against moisture and that the ambient temperature of the location is neither less than 5°C, nor more than 

Summary of Contents for B24M

Page 1: ...Range 4 B24M Operating instructions ECOSAFE Date du document 30 04 2019 Version 01 0 Instruction manual For safety cabinet Model B24M ...

Page 2: ...ONTENT 1 Description Use Page 2 2 Technical features Page 2 3 Installation and commissioning Page 2 4 Internal equipments Page 3 5 Storage Page 3 6 Ventilation Page 3 7 Safety controls Page 4 8 Spare parts Page 4 9 Recycling Page 4 10 Warranty Page 5 11 Control sheet for the user Page 6 ...

Page 3: ...r Manual door that open to a full 180 degrees 3 point locking system Shelves adjustable on 3 76 mm centers for versatile storage Leakproof 2 51 mm bottom sump Dual vents with flame arresters 4 self leveling feet Grounding connector External dimensions H 910 x W 890 x D 560 mm Internal dimensions H 710 x W 795 x D 455 mm Storage capacity 90 liters Retention capacity 18 liters Loading capacity 75 KG...

Page 4: ...t the chosen height thanks to the rack and pinion Take care not to exceed the loading capacity per shelf of 75 KG The cabinet also has an integrated retention tank which has a retention capacity of 18 liters The cabinet is provided with polyethylene trays attached to the galvanized steel shelves and a separate polyethylene liner for the bottom sump 5 STORAGE Your safety cabinet is exclusively dest...

Page 5: ...s On the other hand if the ventilation system of your cabinet is not running you have to indicate it by putting an etiquette on the cabinet saying Safety cabinet not connected to the mechanic ventilation etiquette is not provided 7 SAFETY CONTROLS Your safety cabinet is a safety equipment As to guarantee its perfect efficiency you have to check the equipment regularly Daily safety controls Check i...

Page 6: ...nt as they are the only one able to determine the condition Repair of the equipment must be performed in our warehouse The spare parts are under 3 months warranty upon their settlement The warranty cannot be applied in the following cases The equipment has not been stored used or maintained in accordance with the standards by the user In case of a damage of the equipment by the user or a third par...

Page 7: ...y The locking system of the doors works correctly The mechanic ventilation system is running correctly if applicable The ventilation system is not obstructed The air inlet and outlet are clean The standardized pictogram on the door is in good condition and visible The door hinges are lubricated The locking system is lubricated The containment sump is clean Remarks Date By Signature ...