REMOTE
Puts the calibrator in remote mode. From the remote mode the user can still use the
keypad to get back to local unless the command LOCKOUT was entered before
REMOTE. Than the keypad is totally locked out, and the user has to send the LOCAL
command to get back to local operation.
RTD_TYPE
Sets the RTD type. The following is a list of RTD types the way they should be
entered after the command:
PT385_00; PT392_00; PTJIS_00; CUSTOM;
For Example: RTD_TYPE PT385_00 sets RTD type to Pt385-00
RTD_TYPE?
Returns the RTD type.
For Example: If the RTD type is PT385_00, RTD_TYPE? Will return PT00_385.
ST_START
Starts a switch test.
ST_CLOSE?
Returns the pressure that the switch closed at in the current pressure units.
ST_OPEN?
Returns the pressure that the switch opened at in the current pressure units.
ST_DEAD?
Returns deadband of the switch in the current pressure units.
TEMP_UNIT
This command is used to the temperature unit used when measuring temperature.
The first argument indicates which display to apply the change to. The second
argument is the unit, either CEL for Celsius or FAR for fahrenheit.
For example:
To set the temperature unit to fahrenheit on the lower display send TEMP_UNIT
LOWER, FAR.
TEMP_UNIT?
Returns the temperature unit, (CEL or FAR) used when measuring RTDs for each of the 3 displays.
VAL?
Returns the value of any measurement taking place on the upper and lower display.
For example, if the upper display is measuring 5mA, and the lower display is
measuring 0V, then VAL? will return: 5.000000E-03, A, .0, V
ZERO_MEAS
Zeroes the attached pressure module. Enter the zeroing value in PSI after the
command when zeroing an absolute pressure module.
ZERO_MEAS?
Returns the zero offset or the reference value for absolute pressure modules.