55
11.4.4 Calibrator Status
Error Queue
If an error occurs due to invalid input or buffer overflow, its error code is sent to the error
queue. The error code can be read from the queue with the command FAULT?. The error
queue holds 5 error codes. When it is empty, FAULT? returns 0. The error queue is
cleared when power is reset or when the clear command *CLS is entered.
Input Buffer
Calibrator stores all received data in the input buffer. The buffer holds 250 characters. The
characters are processed on a first in, first out basis.
11.5 Remote Commands and Error Codes
The following tables list all commands, and their descriptions, that are accepted by the calibrator.
Table 5: Common Commands
Command
Description
*CLS
(Clear status.) Clears the error queue
*IDN?
Identification query. Returns the manufacturer, model number, and firmware
revision level of the Calibrator
*RST
Resets the calibrator to the power up state
Table 6: Calibrator Commands
Command
Description
CPRT_COEFA
Sets the custom RTD coefficient A
CPRT_COEFA?
Returns the custom RTD coefficient A
CPRT_COEFB
Sets the custom RTD coefficient B
CPRT_COEFB?
Returns the custom RTD coefficient B
CPRT_COEFC
Sets the custom RTD coefficient C
CPRT_COEFC?
Returns the custom RTD coefficient C
CPRT_R0
Sets the custom RTD R0 resistance
CPRT_R0?
Returns the custom RTD R0 resistance
DAMP
Turns Damp on or off.
DAMP?
Returns if DAMP is on/off
DISPLAY
Turns on/off the displays specified in the command
DISPLAY?
Returns which displays are on/off
FAULT?
Returns the most recent error code
ERROR_LOOP
Turns loop power on or off in percent error mode
ERROR_LOOP?
Returns the current state of loop power in error mode
ERROR_MODE
Turns percent error mode on or off
ERROR_MODE?
Returns whether percent error mode is on or off
ERROR_PORT
Set the pressure port for percent error mode
ERROR_PORT?
Returns the pressure port for percent error mode