29
Command
Value Description
Reply
defSenModNotReady
0xFD
defZeroSpanDataInvalid
0xFE
defOWZeroSpanWrFail
0xFF
ResetO2
0x11
Resets the zero and span for %O2 back to
factory values (M=1, b=0, x=0)
defZeroSpanSuccessful
0x00
defSenModNotReady
0xFD
defZeroSpanDataInvalid
0xFE
defOWZeroSpanWrFail
0xFF
ResetNOX
0x12
Resets the zero and span for NOX back to
factory values (M=1, b=0, x=0)
defZeroSpanSuccessful
0x00
defSenModNotReady
0xFD
defZeroSpanDataInvalid
0xFE
defOWZeroSpanWrFail
0xFF
ResetAllFilters
0x15
Resets alpha for recursive average of %O2
and NOX to factory value (alpha=48)
defAlphaOK
0x00
ExpertModeDisable
0x16
This command removes the unit from expert
mode.
None
EnableH2Calc
0x19
Use H2 formula for lambda calculation.
None
DisableH2Calc
0x1A
Use std formula for lambda calculation.
None
EnableIP1Pcomp
0x1B
Compensate IP1 for Pressure.
None
DisableIP1Pcomp
0x1C
Do not compensate IP1 for Pressure.
PCF=1.0
None
ResetDeltaO2Table
0x1D
Set all delta %O2 values to 0.
None
ResetDeltaLambdaTable
0x1E
Set all delta Lambda values to 0.
None
ResetTPDOs
0x1F
Set all TPDOs as delivered from factory.
None
defFastSensorStart
0x20
Use sensor start parameters in 1wire
None
defSlowSensorStart
0x21
Sensor start draws <1A max
None
defFactoryReset
0xDF
Set all EE values to standard
configuration
None
ExpertModeEnable
0xE0
This command places the unit in expert
mode (password required)
None
HtrPidEnable
0xE1
Enable heater PID control
None
HtrPidDisable
0xE2
Disable heater PID control
None
RvsEnable
0xE3
Enable RVS measurement
None
RvsDisable
0xE4
Disable RVS measurement
None
UseCalibratedValues
0xE5
Use calibrated ADC/DAC values
None
UseUnCalibratedValues
0xE6
Use raw uncalibrated ADC/DAC values
None
UseProgrammedFilter
0xE7
Use programmed recursive average filter
constants (std)
None
UseCalibrationFilter
0xE8
Use slow recursive average filter
constants (cal)
None