Instruction Manual MM850481 ed.01 

5 RECOMMENDATIONS 

 

The analyzer should be used in environments where the temperature is between -5°C and +45°C. 

 

When the analysis are completed, before switch the instrument off, remove the probe from the gas pathway and wait 
about 30 seconds cleaning the pneumatic circuit from gas. 

 

Do not use the analyzer with clogged filters or filled with humidity. 

 

Before placing the probe in the case, be sure that the probe, the water trap and the hose are clean and cold. 

 

If you have ordered a double hose sampling probe, insert both connectors to the instrument during the gas analysis 
(pay attention to color correspondence). 

 

To empty the water trap from the condensate water, remove the draining plug. DO NOT OPEN the water trap. DO 
NOT leave the water trap without the draining rubber plug. 

 

For the best efficiency and accuracy, we suggest to calibrate the instrument every year

.  

 
 

WARNING 

 

I

F THE 

EGA4 

IS STORED AT TEMPERATURE EXCEEDING THE OPERATIVE LIMITS

THE ANALYZER NEEDS SOME MINUTES TO WARM

-

UP 

TO THE ACTUAL AMBIENT TEMPERATURE

BEFORE STARTING THE OPERATION

5.1 Power 

supply 

The instrument can be powered from: 

 

internal rechargeable Ni-MH battery. 

 

External battery charger, supplied as a standard accessory (the batteries must be installed). 

 

Optional external DC auto battery charger (the batteries must be installed). 

 
The Ni-MH rechargeable batteries allows a long time operation and do not need maintenance. The same batteries power 
both the instrument and the internal printer. The internal battery will grant 5 hours continuous operation (without printing 
and display back light off). 
 

N

OTE

: 90% 

BATTERY CHARGING CAN BE OBTEINED IN 

HOURS WITH THE INSTRUMENT SWITCHED OFF

 
During operation a fully battery symbol “

” will be displayed on the display. This symbol means that the batteries are 

completely full. When the batteries will be discharged the symbol “

” will appear and the instrument still has about 20 

minutes operation capability to end the running analysis. 
The battery symbol indicates that a full charge is required. Use only the dedicated battery charger supplied by Eclipse 
together the instrument. 
 

CAUTION: O

LD BATTERIES CAN LEAK AND CAUSE CORROSION

. N

EVER LEAVE RUN DOWN BATTERIES IN THE INSTRUMENT

 

 
 

WARNING 

 

T

HE INSTRUMENT IS SHIPPED WITH AN AVERAGE LEVEL OF BATTERY CHARGE

. A

FTER UNPACKING

A FULL CHARGE OF THE BATTERY 

IS RECOMMENDED

BY CONNECTING THE INSTRUMENT TO THE MAIN LINE THROUGH THE BATTERY CHARGER 

(OFF 

CONDITION

FOR 

8/10 

HOURS

 
 
 

15 

Summary of Contents for EGA4

Page 1: ...EGA4 Combustion Analyzer Instruction Manual 924 2 11 1 03...

Page 2: ...igital technologies Any maintenance operation must be carried out by qualified personnel ONLY Eclipse supplies instructions and operative procedures for any operation on the instrument We recommend to...

Page 3: ...ors 12 4 2 Auxiliary measurements 13 4 2 1 Temperature measurements 13 4 2 2 Pressure and Draft 13 4 2 3 Smoke index measurement 13 4 2 3 1 Measuring instruments option 14 4 2 4 Other measurements 14...

Page 4: ...ation 37 8 5 Configuration Software 38 8 5 1 How operate using GasConfig 38 8 5 2 GasConfig Installation 39 9 CONFIGURATION 40 9 1 Engineering Units 40 9 2 Alarms 40 9 3 HELP key 41 9 3 1 Header setti...

Page 5: ...unit NO NOx and SO2 gas concentration The instruments are completed with a pressure draught sensor an internal printer an internal memory for storing data and a RS232 serial interface for configuratio...

Page 6: ...Serial interface Bi directional infrared interface External IR RS232 converter available for PC communication Synchronous serial port for external keyboard and optical bar pen reader Smoke measuremen...

Page 7: ...ted from 0 to 100 0 1 Tair Pt100 from 10 to 100 C 0 1 C 0 2 rdg 0 15 C Tgas Tc K from 10 to 1000 C 0 1 C 0 3 rdg 0 3 C T Calculated from 0 to 1000 C 0 1 C Tflow Tc K from 10 to 100 C 0 1 C 0 3 rdg 0 3...

Page 8: ...e filter 2 4 Measuring cells EGA4 uses long life sensors for O2 CO H2 compensated NO NO2 and SO2 An automatic device exclude automatically the CO sensor when CO high concentration level is identified...

Page 9: ...it according to the OSHA recommendation are exceeded 2 12 Natural gas probe Optional probe to monitor CH4 and natural gas concentration and detect and locate the position of a gas leak into the domest...

Page 10: ...t Battery charger External probe input Back light sensor Pt100 Tair probe At the bottom of EGA4 you can see all sampling probe connectors gas inlet pressure and draught input thermocouple input for ga...

Page 11: ...s with time The expected life does not relate to the operative time and is lost after approximately 24 months The toxic gas measurements CO SO2 NO and NO2 use electrochemical cells with an expected op...

Page 12: ...RUBBER PLUG THE MEASUREMENTS WILL BE INCORRECT WHEN YOU FINISH THE ANALYSIS REMEMBER TO DRAIN THE WATER TRAP BEFORE PUTING THE GAS PROBE IN THE CARRYING CASE REMUVE THE CONDENSATE WATER FROM THE PROBE...

Page 13: ...The same Tc K input can be used to measure the flow and return temperature in water circuit If you have to measure the water temperature measuring the pipe temperature use an arch model thermocouple...

Page 14: ...y scale sheet with 10 different areas numbered from 0 to 9 The number 0 corresponds to a 85 2 5 reflection Every area reflects 10 less than the previous one The scale is used to compare the scale with...

Page 15: ...om internal rechargeable Ni MH battery External battery charger supplied as a standard accessory the batteries must be installed Optional external DC auto battery charger the batteries must be install...

Page 16: ...input Tgas input IR serial interface RS232 External probes input Battery charger Pt100 Tair input 6 2 Gas probe positioning The measurement site should be arranged at a distance of 2xD behind the exha...

Page 17: ...ST REACH THE FLUE GAS AS A RESULT OF NON GAS TIGHT CONNECTION POINTS BETWEEN THE HEAT PRODUCER AND THE MEASUREMENT SITE PROBE THE FLUE GAS AND EXHAUST GAS PATHWAYS MUST BE CHECKED FOR GAS TIGHTNESS BE...

Page 18: ...y S T W and X They move the cursor symbol or to increase or decrease numerical values ENTER Accept and memory store the modified parameter or variable PRINT Print the report of the analysis ON OFF Swi...

Page 19: ...p NOx 500p Per 10 5 Ref O2 GRAPH Two boxes are positioned on the bottom of the display They contains the name of the function assigned to the F1 and F2 function keys The status bar uses graphical symb...

Page 20: ...ntify the selected character you can move it using the W X S and T keys The box below the characters table shows the currently inserted text The symbol W is the end of line symbol Press CONFIRM to wri...

Page 21: ...ate paragraphs for operations When autozero procedure finished an intermittent sound advise the operator Press the ENTER key to mute the unit The following page will be displayed Now insert probe in c...

Page 22: ...lled the instrument can display the measured values as bar graph Press the GRAPHIC F2 key to change mode O2 CO Ref O2 NUMERIC Using GasConfig Software you can program the graph page Assign up to 6 dif...

Page 23: ...ioning and flue gas intercepting Wait for measurement stabilization Press ENTER to memory store the pressure and draft values Press ESCAPE F1 to return to the previous page Now you can proceed with th...

Page 24: ...asure the true combustion air temperature to have a correct boiler efficiency computation The EGA4 has a auxiliary Rtd input to measure simultaneously the air temperature A Pt100 with a 6 meters long...

Page 25: ...ctor and the terminal the P2 the instrument connector Press MENU to display the following page MENU MEASURE RECALL MEMORY FLOW RETURN TEMP GAS VEL ESCAPE Use the S and T keys to move the cursor symbol...

Page 26: ...S and T keys to move the cursor symbol to NO copy or Model field Press the ENTER key to change the number of the reports you have to print and or the print model for the report Press the PRINT key aga...

Page 27: ...eys to move the cursor symbol to MEASURE option Press the ENTER key to select the option Press ESCAPE F2 to return to the previous menu MEASURE P DECAY LEAK TEST GAS LEAK TEST AMBIENT CO ENVIRON TEMP...

Page 28: ...e cursor symbol to the value to be changed Press the ENTER key to store the new selection Select START TEST option and press the ENTER key to run the decay leak test P1 100 0 mmH2O P2 99 5 mmH2O P3 __...

Page 29: ...Press ESCAPE F2 to return to the previous menu MEASURE P DECAY LEAK TEST GAS LEAK TEST AMBIENT CO ENVIRON TEMP HUM ESCAPE Use the S and T keys to move the cursor symbol to AMBIENT CO option Press the...

Page 30: ...F PROBE IS CONNECTED IN GAS LEAK TEST PAGE A WAITING MESSAGE WILL BE DISPLAYED Refer to paragraph 10 2 3 to change the 60 sec heating time Before insert the probe in working area agitate it for same s...

Page 31: ...ENVIRON TEMP HUM ESCAPE Use the S and T keys to move the cursor symbol to ENVIRON TEMP HUM option Press the ENTER key to select the option Press ESCAPE F2 to return to the previous menu Connect the pr...

Page 32: ...T SELECT THE TAG IF IT IS DIFFERENT FROM THE CURRENT 8 1 Store data analysis If you had pressed the ENTER key with the cursor on the STORE option the analyzer will display the following message WAIT s...

Page 33: ...d boiler information to select up to 40 boiler to be check to assign an operator name and transfer the data to the analyzer The operator will memory store up to 250 boiler checks and print the report...

Page 34: ...ect in automatic mode all analysis made by more than XX months Pressing the Show selected button the program will show all boiler selected for data transfer You can exclude manually one or more boiler...

Page 35: ...boiler information for the selected plant and analysis groups data for the selected boiler The analysis groups data will be ordered by data and will show the general data for the group of measurements...

Page 36: ...cs Select the column by positioning the mouse pointer on the desired column header and click using the mouse left button click using the mouse right button and select the Select for graph option The c...

Page 37: ...16Mbyte Ram and 20 Mbyte on hard disk a color monitor and a Microsoft mouse or a compatible one In order to install DBGas 2000 for Windows follow the below procedure Place the DBGas 2000 disk in a CD...

Page 38: ...epends by instrument model installed cells and options The window on the right is the program user interface It is organized in function pages you can select the page for interesting parameters using...

Page 39: ...d fax numbers 8 5 2 GasConfig Installation Software runs on IBM PC under WINDOWS 95 98 NT operative systems Minimum requirements are a Pentium CPU with 16Mbyte Ram and 8 Mbytes on hard disk a color mo...

Page 40: ...ation software you can program the CO level for automatic exclusion the CO cell Fuel Parameters You can display only all the constant table for the selected fuel To change the values you should use th...

Page 41: ...the beep will stop but the reverse mode remain 9 3 HELP key Help menu inform of the instrument status electrochemical cells status and battery level Service information and allows you to change the h...

Page 42: ...information to contact us for service assistance and the information that we wont to know for identify your instrument configurations instrument configuration code and serial number firmware version...

Page 43: ...eference damaged Return the instrument to Eclipse for maintenance Over Overrange condition The problem is solved automatically when measure return inside the range Under Underrange condition The probl...

Page 44: ...10 2 1 Clock Select this option for date and time set up Verify that the current values are corrects because the analyzer use these data for analysis storing CLOCK Hour 10 Minutes 22 ESCAPE DATE Use t...

Page 45: ...SIS AutoZ Time 180 CH4 Heat Time 60 ESCAPE Use the S and T keys to move the cursor symbol to the desired option Press the ENTER key to set a new Auto zero time Press ESCAPE F2 to return to the previou...

Page 46: ...key while pressing the ON OFF key in order to set the instrument in firmware upgrade status 9 Click on the Program button to start the upgrading phase and wait until a termination message appears NOT...

Page 47: ...shows evidence of being damaged as a result of excessive corrosion or current heat moisture or vibration improper specification misapplication misuse or other operating conditions outside of Eclipse...

Page 48: ...L Natural Gas TAG SELECTION ANALYS 3 Connect the sampling probe electrical and pneumatic terminals to the EGA4 4 Insert the sampling probe in chimney Be sure for mechanical positioning and flue gas in...

Page 49: ...5 Firmware Upgrade 46 Flow Return temperature 24 Fuel technical data 8 Fuels menu 36 G Gas analysis 21 Gas leak cheker 30 Gas probe positioning 16 Gas sampling probe 8 11 Gas sensors 12 GENERAL 8 GEN...

Page 50: ...a 42 Set up menu 43 Smoke index measurement 13 Smoke measurement 9 23 Specifications 6 status bar 19 Store data analysis 32 T T RH probe 9 Temperature Humidity probe 30 Temperature measurements 13 Tem...

Reviews: