Ecleris PROCAM HD Manual Download Page 11

INMANUUIN121 M. USO VI340-HC118G2 ING USA V.2016.10.06.DOCX                   8 

 

 

Once the coupling is finished, the combined set ECLERIS Everlux light source and PROCAM HD console is ready to use as itself or it can be attached 
to  an  ECLERIS  microscope  forearm  or  to  an  ECLERIS  colposcope  forearm  through  the  ECLERIS  Everlux  light  source  (after  removing  the  Everlux 
handle on top of the unit). 

 

In order to avoid both the use of two electrical power sockets to supply both devices and also the use of many cables, ECLERIS has specially designed 
an  interlock  male-female  cable  of  24  centimeters.  With  this  cable,  the  PROCAM  HD  console  can  be  supplied  by  the  ECLERIS  Everlux  light  source 
directly. This cable is optional and it has to be requested by the user. 

This cable connects both units as it is shown in the picture. 

Summary of Contents for PROCAM HD

Page 1: ...INMANUUIN121 M USO VI340 HC118G2 ING USA V 2016 10 06 DOCX PROCAMHD CCDCAMERA AND CONSOLE VI340 HC118...

Page 2: ...tive MEDES Ltd UK 5 Beaumont Gate Shenley Hill Radlett Herts WD7 7AR England Tel 423 663 169205 Tel Fax 44 1923 859 810 e mail medes arazygroup com Ecleris PROCAM cameras are CE approved complying wit...

Page 3: ...2 PROCAM HD Camera 5 4 INSTALLATION 5 4 1 COUPLING TO ECLERIS EVERLUX LIGHT SOURCE 6 5 REVISION AND OPERATION 9 5 1 REVISION 9 5 2 OPERATION 9 5 2 1 White balance 10 5 2 2 Window Function 10 5 2 3 Fu...

Page 4: ...up Enviroment transport conditions Pressure Limits Humidity Limits Camera connection Window Function White Balance Anti Moir Function CE Mark Manufacturer Information Name Address Type BF Do not disc...

Page 5: ...s proper electrical isolation If it needs to be replaced the replacement will not be longer that the original cable and it needs to be certified under IEC 60245 1 or IEC 60227 1 In case of interferenc...

Page 6: ...ng flexible fiber endoscopes and Windows minimizes the light saturation when using small diameter endoscopes PROCAM HD power supply console was specifically designed to be used together with High Defi...

Page 7: ...X 4 3 3 Components 3 3 1 PROCAM HD Console 3 3 1 1 Front Panel 1 ON OFF button 2 PROCAM HD Connector 3 3 1 2 Rear Panel 1 Power connector 2 Equal potential Ground connection 3 Video output in high def...

Page 8: ...or later Use the endoscope in accordance with IEC60601 2 18 1 Unpack the device with care 2 Place the PROCAM HD console on a solid surface 3 If the building structure has equipotential grounding conne...

Page 9: ...x light source you will need a 2 5 millimeters Allen key With this Allen key unscrew the two screws at the bottom of the PROCAM HD console as it is shown in the picture Doing so two heads of screws wi...

Page 10: ...in place into the holes of the ECLERIS Everlux light source base you need to slightly push back the PROCAM HD console in order to attach it completely to the ECLERIS Everlux light source Finally to e...

Page 11: ...Everlux light source after removing the Everlux handle on top of the unit In order to avoid both the use of two electrical power sockets to supply both devices and also the use of many cables ECLERIS...

Page 12: ...to the electrical socket through the power cable supplied by ECLERIS The electrical supply has to be done at electrical socket without power adapters The equipment must be used only in places where t...

Page 13: ...ctivated only the surface that contains image information is used to control the camera sensitivity this way the camera adjustment and the image contrast is optimized even for very thin optics and sho...

Page 14: ...r Turn off Anti Moir Deactivated Blinking Anti Moir level 1 Turn on Anti Moir level 2 To active the first level of filtration press the Anti Moir button on the camera once When the first level is acti...

Page 15: ...ially alcohol when the unit is hot Do not turn the equipment on until one 1 hour after cleaning process We recommend cleaning the external surface of the equipment every 15 days Do not sterilize any p...

Page 16: ...NMANUUIN121 M USO VI340 HC118G2 ING USA V 2016 10 06 DOCX 13 Only use fuses as shown at the rear side of the PROCAM HD console and certified by IEC60127 4 Screw again the knobs 5 Connect the power cor...

Page 17: ...ION 1920 H x 1080 V 60 fps SENSOR CMOS THERMAL PROTECTION Overheating automatic shutdown of power supply OPERATING CONDITIONS 10 40 C 35 75 relative humidity 700 1060 hPa STORAGE AND TRANSPORT 30 45 C...

Page 18: ...nput change the connection to another input Monitor off Turn the monitor on Monitor disconnected Plug the monitor to the PROCAM HD console through the provided cord Monitor in wrong video mode Set the...

Page 19: ...nsible for damage caused during the transportation due to improper packaging 10 WARRANTY ECLERIS declares that the PROCAM HD has passed our Quality Control and no defects have been found in its manufa...