4.  INSTALLATION AND START UP 

4.1. INSTALLATION 

Experienced personnel are not required for the installation of the Ecleris Minivac. 

1)  Place the device on a flat surface. 
2)  Check that the plastic collecting bottle is firmly placed. Matching shall be perfect and there 

shall be no air leakage. Check that the micro-porous filter is installed inside the bottle. 

3)  Check that the ventilation grids are free of obstructions. 
4)  Connect the suction hose (1.5 meter quick assembly coupling hose) to the device. 
5)  Connect the knurled hand piece (1) to the free end of the suction hose, pressing both parts. 

There should be no air penetration and both parts should fit perfectly. 

6)  Screw the vacuum regulator (2) to the distal end of the knurled hand piece (1). 
7)  The  45º  angular  (3)  adapter  can  be  placed  in  the  free  end  of  the  vacuum  regulator 

depending on the use: perpendicular to the tissue or at 45º. Place a 50, 75, 100, 150 or 200 
microns Diamond Tip depending on the exfoliation depth desired. 

 

 

 

 

 

 

 

 

 

O-RING COLOR 

DIAMOND GRADE 

White 

50 

Yellow 

75 

Orange 

100 

Green 

150 

Purple 

200 

 

8)  Connect  the  AC  interlock  cable  to  the  Ecleris  Minivac.  Check  that  the  electrical  network 

voltage to be used is the same as the one stated in the voltage selector switch on rear panel 
of the device and that the correct fuse is installed. 

4.2. REVISION 

Once  the  installation  is  finished,  and  before  starting-up  the  device,  please  goes  through  the 
following checklist: 

  Check that the electrical network voltage to be used is the same as the one stated in the 

voltage selector switch on rear panel of the device. 

  Verify that the electrical network used has its grounding connection in perfect conditions, 

and that it makes perfect contact with the device. 

Summary of Contents for MINIVAC

Page 1: ...ny electronic data processing form The information contained in this manual may be modified or broaden without previous notification without it representing any responsibility to ECLERIS ECLERIS S R L...

Page 2: ...UP 7 4 1 INSTALLATION 7 4 2 REVISION 7 4 3 VACUUM SYSTEM VERIFICATION 8 5 OPERATION 8 5 1 VACUUM INTENSITY 8 5 2 DIAMOND TIP 9 5 3 PRE TREATMENT RECOMMENDATIONS 9 5 4 TREATMENT 9 5 5 POST TREATMENT RE...

Page 3: ...ontact your local distributor or the nearest ECLERIS office ECLERIS shall not be liable for the damages that may occur if the equipment is not used according to these instructions and following the cr...

Page 4: ...e manufacturer specific intent use will exclude ECLERIS of all liability This user manual must be with the device for permanent review If the system is sold or donated the user manual must be delivere...

Page 5: ...in disorders and those caused by aging Due to microdermabrasion the skin becomes more receptive and the absorption of applied products increases without causing any pH alteration The Ecleris Minivac p...

Page 6: ...proceed with the installation 1 Main unit 1 AC interlock cable 1 Quick assembly coupling hose 1 Diamond Tip Kit 1 User s manual Contact your supplier for customized Diamond Tip kits availability 3 4...

Page 7: ...knurled hand piece 1 7 The 45 angular 3 adapter can be placed in the free end of the vacuum regulator depending on the use perpendicular to the tissue or at 45 Place a 50 75 100 150 or 200 microns Dia...

Page 8: ...lating valve clockwise all the way 2 Place the applicator on the back of your hand As soon as it gets in contact with your skin suction will begin The maximum vacuum level should be approximately 60cm...

Page 9: ...inest tip and with low vacuum pressure Once the three parameters have been decided regulate vacuum intensity Vacuum pressure controls the skin pressure on the diamond tips 50 and 75 micron tips and a...

Page 10: ...aning the Ecleris Minivac disconnect the power supply cable from the electrical network 6 1 1 EXTERNAL SURFACES Clean the external surface of the equipment using a piece of cotton or cloth moistened w...

Page 11: ...s they will produce air leakage lowering the vacuum level Never use the Ecleris Minivac without the micro porous filter and the collecting receptacle The Ecleris Minivac was not designed for liquid ab...

Page 12: ...FILTER Micro porous Micro porous FUSES T5AL250V T2 5AL250V T4AL250V T2AL250V CONSUMO CONSUMPTION 275 VA 90 VA DIMENSIONS 35x22x16 cm 35x22x16 cm WEIGHT 4 Kg 5 5 Kg OPERATING CONDITIONS 10 40 C 35 75 r...

Page 13: ...again The Ecleris Minivac starts but does not produce vacuum The vacuum regulator is completely open Close the regulator by turning it clockwise The suction piping is throttled Check that the hose is...

Page 14: ...ame Owner s address Person to contact and a telephone number Product serial number Problem description The shipment should be done in the original packaging to avoid damage to the equipment during the...

Reviews: