4.2 MOUNTING THE SET TO THE GANTRY 

1)  Loose 

the 3 screws from the column’s free side with a 3/16” hex key. 

2) 

Install the forearm introducing the forearm’s coupling into the column’s receptacle. 

3)  Adjust the 3 screws firmly. 

 

4.3 MOUNTING THE CONSOLE 

1) 

Loose the console’s fixation supports, located in the forearm. Use the 4 mm hex key. 

2)  Put  the  light  source  below  the  forearm  matching  the  supports  with  the  orifices  of  the 

console’s front face. 

3) 

Move the console upwards so that the supports are introduced into the console’s orifices. 

4)  Move  the  console  backwards  so  that  the  supports  slide  in  the  guides  and  support  the 

console. 

5)  Ajuste los tornillos de los soportes, de modo que la consola quede sujeta en forma segura 

al antebrazo. 

 

6)  Connect  the  cables  that  come  out  from  the  arm  to  the  respective  connectors  in  the 

console’s panel. 

7)  Connect the power supply cord to the alternating current socket and connect the cord to 

the mains electricity. 

Summary of Contents for C100HD

Page 1: ...without previous written authorization from ECLERIS Neither can it be reproduced multiplied or published through any electronic data processing form The information contained in this manual may be mod...

Page 2: ...BASE 6 4 2 MOUNTING THE SET TO THE GANTRY 7 4 3 MOUNTING THE CONSOLE 7 4 4 THE GANTRY S ADJUSTMENTS 8 5 CONTROLS 9 5 1 HEAD 9 5 2 THE CONSOLE S FRONT 9 5 3 THE CONSOLE S BACK PANEL 10 6 OPERATION 11...

Page 3: ...ould be used in offices where the electrical installation is grounded pursuant to the relevant rulings for health facilities The HD VIDEO COLPOSCOPE should be connected to the mains electricity throug...

Page 4: ...vice during electrical storms Do not introduce objects into the orifices and do not place containers with fluids on the device Disconnect the power supply cable from the outlet when the device is not...

Page 5: ...hic studies in vivo and measuring the size of the lesions with capacity for diagnosing a multiplicity of them The Video Colposcopy study allows identifying the subtle characteristics of the epithelium...

Page 6: ...irm and horizontal surfaces with an inclination of no more than 5 Avoid exposing the device and its accessories to direct sunlight dust or excessive vibrations or bumps Remove the device s parts from...

Page 7: ...light source below the forearm matching the supports with the orifices of the console s front face 3 Move the console upwards so that the supports are introduced into the console s orifices 4 Move th...

Page 8: ...mechanical brakes to regulate the arm and the forearm s different positons as well as regulating the head s height so as to ensure a stable focus and the operator s comfort The head also has a brake...

Page 9: ...may be used with the light beam so as to have a special lighting on the tissues This increases the contrast of the vascular beds and provides a better histological differentiation on the basis of the...

Page 10: ...back panel hosts the power supply connector with fuse holder the equipotential ground terminal which should be duly connected to the equipotential ground node of the medical facility and two HD video...

Page 11: ...y with the light off the user may control it manually using the joystick F F White balance The white balance is set from the factory and it is stored in the device s non volatile memory However if it...

Page 12: ...NU In order to access to the advanced menu press any key on the device s head while it is turning on while the three head s lights are flickering for approximately 10 seconds Use the joystick s comman...

Page 13: ...ions Revision of the electrical safety parameters 7 1 CLEANING Before cleaning the HD colposcope disconnect the cord from the mains electricity Do not sterilize or submerge the device or its accessori...

Page 14: ...cover to replace the damaged fuse 3 Remove the damaged fuse and replace it with another one 4 Use the fuses pursuant to the marks on the device s labels 5 Introduce the socket again in the same positi...

Page 15: ...240 300 mm minimum magnification Field of view 90x50 mm 300 mm with minimum magnification Magnification Zoom levels depend on the monitor size in the following table some of the possible monitor size...

Page 16: ...eo cable is disconnected Check the connection between the video cable and the console s front panel The head does not emit light The video connector is disconnected Check the connection between the vi...

Page 17: ...tion for the distance from the target Decrease the magnification level See zoom level The central zone of the image has multiple focal distances Deactivate the autofocus and perform the focus adjustme...

Page 18: ...ect in the manufacture of same Therefore the appropriate service under the applicable conditions of use indicated herein is guaranteed The warranty period is of one 1 year effective as of the date of...

Reviews: