Ecleris C100HD User Manual Download Page 3

USER MANUAL

– C100HD 

 

Page 

3

 of 

18

 

INMANUUIN165 Rev.2019.04.01 / ENG 

1.  SYMBOLS 

 

Warning  

 

EU Authorized 
Representative 

 

Read the related documentation. 

 

This side up 

 

Alternate current 

 

Fuse 

 

Applicable part type BF 

 

Video output 

 

Protective Earth (Ground) 

 

Freeze image 

 

Equipotential bonding 

 

Video camera 

 

Manufacturer  Information  (Name, 
Address) 

 

White balance 

  

 

 
Serial Number 

 

Lighting 

 

Date of Manufacture (MM/AAAA) 

 

CE mark 

 

Do  not  look  straight  to  the  light 
source 

 

Temperature limits 

 

Do not disassemble 

 

Pressure Limits 

 

Do not discard with the household 
waste 

 

Humidity Limits 

 

Fragile 

 

Keep dry and protected 
from moisture 

2.  SAFETY INFORMATION 

2.1 WARNINGS 

  In  order  to  avoid  electrical  shock,  this  device  should  only  be  connected  to  a  grounded 

outlet 

  Do not install the device near sources of heat such as heating 

  The HD VIDEO COLPOSCOPE should be used in offices where the electrical installation 

is grounded pursuant to the relevant rulings for health facilities 

  The HD VIDEO COLPOSCOPE should be connected to the mains electricity through the 

cord supplied or a replacement of same with the same certifications.  

  The mains electricity outlet should be grounded and no adaptor should be used. 

  The  sectionalizing  of  the  electricity  should  be  performed  with  an  electrical  switch  that 

meets  the  IEC  61058-1  standard,  the  guaranteed  sectionalizing  should  be  performed 
either disconnecting the console’s power cable or removing the coupler from the outlet. 

  The  device  is  grounded  with  a  ground  conductor  on  the  power  supply  cord.  The  ground 

connection is essential for the safe operation of the device. 

  The device is connected to an equipotential bond through a ground terminal on the power 

supply cable. This grounding is essential for the safe operation of the device. 

  Do  not  connect  the  device  to  visualization  devices  that  do  not  meet  the  standard 

IEC60601-1. 

Summary of Contents for C100HD

Page 1: ...without previous written authorization from ECLERIS Neither can it be reproduced multiplied or published through any electronic data processing form The information contained in this manual may be mod...

Page 2: ...BASE 6 4 2 MOUNTING THE SET TO THE GANTRY 7 4 3 MOUNTING THE CONSOLE 7 4 4 THE GANTRY S ADJUSTMENTS 8 5 CONTROLS 9 5 1 HEAD 9 5 2 THE CONSOLE S FRONT 9 5 3 THE CONSOLE S BACK PANEL 10 6 OPERATION 11...

Page 3: ...ould be used in offices where the electrical installation is grounded pursuant to the relevant rulings for health facilities The HD VIDEO COLPOSCOPE should be connected to the mains electricity throug...

Page 4: ...vice during electrical storms Do not introduce objects into the orifices and do not place containers with fluids on the device Disconnect the power supply cable from the outlet when the device is not...

Page 5: ...hic studies in vivo and measuring the size of the lesions with capacity for diagnosing a multiplicity of them The Video Colposcopy study allows identifying the subtle characteristics of the epithelium...

Page 6: ...irm and horizontal surfaces with an inclination of no more than 5 Avoid exposing the device and its accessories to direct sunlight dust or excessive vibrations or bumps Remove the device s parts from...

Page 7: ...light source below the forearm matching the supports with the orifices of the console s front face 3 Move the console upwards so that the supports are introduced into the console s orifices 4 Move th...

Page 8: ...mechanical brakes to regulate the arm and the forearm s different positons as well as regulating the head s height so as to ensure a stable focus and the operator s comfort The head also has a brake...

Page 9: ...may be used with the light beam so as to have a special lighting on the tissues This increases the contrast of the vascular beds and provides a better histological differentiation on the basis of the...

Page 10: ...back panel hosts the power supply connector with fuse holder the equipotential ground terminal which should be duly connected to the equipotential ground node of the medical facility and two HD video...

Page 11: ...y with the light off the user may control it manually using the joystick F F White balance The white balance is set from the factory and it is stored in the device s non volatile memory However if it...

Page 12: ...NU In order to access to the advanced menu press any key on the device s head while it is turning on while the three head s lights are flickering for approximately 10 seconds Use the joystick s comman...

Page 13: ...ions Revision of the electrical safety parameters 7 1 CLEANING Before cleaning the HD colposcope disconnect the cord from the mains electricity Do not sterilize or submerge the device or its accessori...

Page 14: ...cover to replace the damaged fuse 3 Remove the damaged fuse and replace it with another one 4 Use the fuses pursuant to the marks on the device s labels 5 Introduce the socket again in the same positi...

Page 15: ...240 300 mm minimum magnification Field of view 90x50 mm 300 mm with minimum magnification Magnification Zoom levels depend on the monitor size in the following table some of the possible monitor size...

Page 16: ...eo cable is disconnected Check the connection between the video cable and the console s front panel The head does not emit light The video connector is disconnected Check the connection between the vi...

Page 17: ...tion for the distance from the target Decrease the magnification level See zoom level The central zone of the image has multiple focal distances Deactivate the autofocus and perform the focus adjustme...

Page 18: ...ect in the manufacture of same Therefore the appropriate service under the applicable conditions of use indicated herein is guaranteed The warranty period is of one 1 year effective as of the date of...