60
Net weight 4.5KGS
Gross weight 6.0KGS
C.3 Environment Conditions
z
Transportation
Temperature -20
℃~
+50
℃
Relative humidity 25%
~
95% (Non-condensing
)
Atmospheric pressure 500Pa
~
1060Pa
z
Storage
Temperature -20
℃~
+50
℃
Relative humidity 25%
~
85%
Atmospheric pressure
500Pa
~
1060Pa
z
Usage
Temperature +5
℃~
+40
℃
Relative humidity 25%-80%
Atmospheric pressure 700Pa
~
1060Pa
Appendix D EMC information
EMC guidance and manufacturer’s declarations - Electromagnetic emissions -For all the
equipments and systems
EMC guidance and manufacturer’s declarations - Electromagnetic emissions
EM-1201/1202/1203 ECG Machine are intended for use in the electromagnetic environment specified
below. The customer or user of EM-1203/1202 electrocardiograph should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
EM-1201/1202/1203 ECG Machine use RF energy only
for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
EM-1201/1202/1203 ECG Machine are suitable for use
in all establishments, including domestic establishments
and those directly connected to the public low voltage
power supply network that supplies buildings used for
domestic purposes.