29
Identifier
Name
Description
A
Operation Mode
Select Rhythm, AUTO and Manual mode from the keyboard
B
Baseline Drift Filter
Select it in the filter setup interface, refer to Chapter 5.9.7
C
EMG filter
Select it in the filter setup interface, refer to Chapter 5.9.7
D
Time and date
Set it in the time and date setup interface, refer to Chapter 5.9
E
Power supply mode
Indicate the current power supply mode, AC or battery
F
Heart rate
Current Heart rate value
G
Name
Patient name: no more than 20 characters
H
ID
Patient ID: no more than 10 digits
I
Leads and waveform
display
Display the waveforms
F1 Patient
Press
【
F1
】
, patient information input
F2
Paper speed setup
Press
【
F2
】
, choose paper speed within 5mm/s, 6.25mm/s,
10mm/s, 12.5mm/s,25mm/s,50 mm/s
F3 Sensitivity
setup
Press
【
F3
】
, choose sensitivity within 2.5mm/mV. 5mm/
mV,10mm/ mV,20mm/mV
F4 Freeze
Press
【
F4
】
, save current patient test waveforms
F5 Lead
Press
【
F5
】
to see electrodes placement indication
F6 Setup
Press
【
F6
】
to enter system setup interface, refer to chapter 5.9
F7 More
Press
【
F7
】
,enter the second page of the main interface
Replay
Press
【
F4
】
to Freeze the waveform displayed on the main interface.