Ecel CURING LIGHT EC 500 Owner'S Manual Download Page 6

Discharged battery:

Exhausted battery:

Preventive maintenance:

.

Nature of Preventive maintenance:

Circuit diagrams, list of parts, components, and others:

Curing Light EC 500

the equipment emits two tones consecutives every 10 seconds of

operation of the Led.

the equipment emits three tones consecutives and turns the Led off,

preventing the operation.
Perform the procedure to recharge the battery if the equipment indicates Discharged
battery or Exhausted battery.

Clean the equipment surface with a clean soft cloth moistened with neutral detergent. The
light guide tip must be cleaned with gauze or soft cloth.

Never involve the acrylic tip with any other kind of material.
Do not use a solution containing alcohol, solvents, ammonia or abrasives.
Do not use any instrument or abrasives on the sides of the top of the handpiece,
because it can damage it, causing loss of light emission.
The equipment must be disconnected from the electric outlet before cleaning.

In case the equipment presents any faults that are not listed in the “Troubleshooting”
section of this manual, contact the Authorized Technical Assistance. Do not open the
product and/or try to repair it. This action may aggravate the problem or even generate
further malfunctions.

In order to assure the electrical security of the equipment during

its useful life, we recommend that it should be verified by Authorized Technical Assistance at
least one time each two years. (After warranty)

Should be checked the following items: light intensity,

cracks and dryness in the power cable, tension in the charger, the tension of the battery
charger, time settings and general state of the light guide tip.

The supply of circuit diagrams,

list of parts, components or other necessary information for Technical Assistance, can be
requested from the manufacturer and will be supplied by agreement.

This equipment must be used only by properly trained personnel under supervision of
qualified dental staff.
It is advisable that the professional informs his patients or any other individuals who have
access to the equipment about the cautions to be taken during its handling.
This equipment is not appropriate for use in areas at risk of explosion.
Before the first use and/or after long work interruptions, such as vacation periods, clean and
disinfect the equipment.
Do not expose it directly to sunlight, because it may cause early wear.
Before connecting the charger to the power supply, make sure that the main voltage is in
accordance with the charger's voltage range.
Do not look at the light without the eye protector or the goggle, because it may be harmful.
Before using the equipment, wait until the unit reaches the room temperature, especially
when it is taken from cold conditions into warmer ones.
Do not use another battery except the original of

. Otherwise, your

equipment may be denied.
Do not place the unit near flammable materials, and do not touch the flammable solutions
with the light.
Do not touch the battery contacts with metal parts, as this may damage it permanently.

6

Recommendations, Precautions and Warnings

!!

Maintenance

Cleaning and Care

Summary of Contents for CURING LIGHT EC 500

Page 1: ...Owner s Manual CURING LIGHT EC 500 ...

Page 2: ...IV O LT AG EM E 00 00 00 EC EL In dú st ria e Co m ér ci o Lt da Av M og ia na 18 54 Ri be irã o Pr et o SP Br as il CE P 14 07 5 26 0 PO TÊ NC IA 10 VA FR EQ 50 60 Hz NÚ ME RO DE SÉ RI E SE RI AL NU MB ER MO DE LO EC 50 0 TE NS ÃO MU LT IV OL TA GE M E 00 00 00 EC EL Ind ús tri a e Co mé rci o Ltd a Av Mo gia na 18 54 Ri be irã o Pr eto SP Br as il CE P 14 07 5 2 60 ...

Page 3: ...090 848 0001 47 Mogiana Avenue 1854 Jd Independência Ribeirão Preto SP CEP 14075 260 Phone 16 3878 1700 Walter Bagatta Junior CREA SP 0601359652 Declared as exempt from registration by the Brazilian Health Department Figures Product Presentation Curing Light EC 500 Installation Procedure Before Product Use Product Description Operation Time Settings Charging the Battery Recharging the Battery Batt...

Page 4: ...the acrylic curing light tip to the front orifice of the handpiece Plug the eye protector in the acrylic tip Set the handpiece on the charger base Find a fresh and dry place far from dust humidity chemicals products and cleaning agents Charge the battery during 12 hours before the first use After attaching the tip to the equipment it should be cleaned with some gauze since oiliness from the user s...

Page 5: ... at this moment the LED will shut down keep pressing the button It will start a sequence of 10 tones with pauses of approximately one second between them Each tone corresponds to 10 seconds of the time setting Release the activation button after getting the desired number of tones which will correspond to the time working When the equipment is inserted in the charging base the battery indicator Le...

Page 6: ...ld be checked the following items light intensity cracks and dryness in the power cable tension in the charger the tension of the battery charger time settings and general state of the light guide tip The supply of circuit diagrams list of parts components or other necessary information for Technical Assistance can be requested from the manufacturer and will be supplied by agreement This equipment...

Page 7: ...quipment on patients with cardiac pacemakers Dispose the battery at an appropriate site according to local regulations avoiding environmental contamination and its inadequate use was designed to do not be sensitive to interferences as electromagnetic fields electric external influences electrostatic discharge or pressure and variations of pressure as long as the equipment be installed cleaned main...

Page 8: ...525mmHg to 795mmHg ENVIRONMENTAL OPERATION CONDITIONS Atmospheric pressure range Environmental temperature functioning range 10 C to 35 C Environmental temperature range recommended by ECEL 21 C to 26 C Functioning relative humidity range non condensing 30 to 75 700hPa to 1060hPa 525mmHg to 795mmHg After Installation Environmental temperature range accommodation 5 C to 45 C Accommodation relative ...

Page 9: ...TATION STORAGE ENVIRONMENTAL CONDITIONS Transportation and storage environmental temperature range 0 C to 55 C Transportation and storage relative humidity range non condensing 0 to 90 Atmospheric pressure range 500hPa to 1060hPa 375 mmHg to 795 mmHg The product must be transported and stored Carefully in order to prevent vibrations and impacts By keeping the side indicated by the arrow facing upw...

Page 10: ...r Installation and or operation in disagreement with the instructions in this manual and falls Technician s travel expenses and equipment transportation expenses for technical assistance External finishing and or accessories damages not caused by the manufacturer Present the sales receipt to theTechnicalAssistance during the warranty Reasons for warranty voidance Items that are not covered by the ...

Page 11: ...e battery again Solution Equipment does not charge Fault Cause Charger is disconnected or is not well connected Solution Verify if the charger is disconnected Equipment does not cure Fault Cause Obstruction of the light guide tip Solution Replace or clean it Cause There is no power supply Solution Await for power supply normalization or call for technical assistance Cause LED is burned out Solutio...

Page 12: ...ECEL ABRIL 2009 Ecel Indústria e Comércio Ltda Fone 55 16 3878 1700 ecel ecel ind br www ecel ind br Av Mogiana 1854 Jd Independência Cep 14075 260 Ribeirão Preto SP Brasil ...

Reviews: