32
3.
Calculate and Record the slope of the electrode, 58.2 mV/pH
4.
Install the electrode into the field mounted sensor
5.
Press MANUAL and enter the pH value, 7.00 pH, press mV and enter the
corresponding mV value, 6.8 mV, press OK, Accept Offset?, press YES, enter
slope 58.2 mV/pH, press OK, Accept Slope?, Press YES
6.
The Calibration is complete, the Offset and Slope values are displayed, press
OK to exit.