39. If the swing-away tray was
ordered follow steps 39-42, using a wrench,
remove the nut from the bolt on the tray arm
and discard the nut. Make sure all other
hardware stays on the bolt.
39
37. Loosen knobs on the tray brackets and
flip open the gray latch all the way as shown
below.
38. Lock the casters, insert the two tray
arms into the tray bracket on the bottom of
the tray as shown. Flip gray latch closed,
and tighten knob securely.
Caution: After placing the tray into position,
ensure the ratchet handles on both sides of
the tray are tightened securely.
37
38
!
Page 15
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The EasyStand Zing is intended for use in the electromagnetic environment specified below. The customer or the user
of the EasyStand Zing should assure that it is used in such an environment.
Immunity Test
Conducted RF
IEC 61000-4-6
6 Vrms
150 kHz to 80
MHz
6 Vrms
Portable and mobile RF communications equipment
should be used no closer to any part of the
EasyStand Zing, including cables, than the recom-
mended separation distance calculated from the
equation applicable to the frequency of the transmit-
ter.
Recommended separation distance
d = 1.2
√
P
d = 1.2
√
P
80 MHz to 800 MHz
d = 2.3
√
P
800 MHz to 2.7 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as deter-
mined by an electromagnetic site survey
a
, should
be less than the compliance level in each frequency
range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
IEC 60601
Test Level
Electromagnetic Environment - Guidance
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
Compliance
Level
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7
GHz
3 V/m
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the EasyStand Zing is used exceeds the
applicable RF compliance level above, the EasyStand Zing should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the
EasyStand Zing.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Note: The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a residential environment (for with CISPR 11 class B is normally
required) this equipment might not offer adequate protection to radio-frequency communication services.
The user might need to take mitigation measures, such as relocating or re-orienting the equipment.
MPS, Supine, TT
MPS, Supine, TT
MPS, Supine, TT
Page 24