Page | 11
before reach power supply)
failure.
correct replace the unit.
Unit is ON but display is not
showing.
not powered on.
Check the power switch is on
position.
Fan does not rotate
Fan wiring is loose
Checking the wiring
Fan motor is fail
Change parts
High temperature in cabinet
The temperature setting of
outage is too high
Set the proper temperature set
point.
Refrigerant leaks
Find leak location and make
repair welding. Infuse refrigerant
again
Temperature probe is damaged
Change parts
Capillary tube or system is
blocked by dirt\ greasy
Clean the capillary tube or
change filter.
High ambient temperature
Turn on air-conditioner
,
reduce
ambient temperature.
Specifications
Model
ULTH-100
Capacity
100 LITER
Rating voltage
230 V
Rating Frequency
50 Hz
Rated current
5.5 A
Compressor 1
st
stage
EMBRACO 2155
Compressor 2
nd
stage
EMBRACO 2140
Refrigerant 1
st
stage
R-404a (350 gm)
Refrigerant 2
nd
stage
R-508b (200 gm)
R-290 (20 gm)
Temperature Controller
PPI ZENEX 44
Cascade Controller
Subzero Sz-7569-P
Voltage Current Protector
Gelco AaCcSvch
Design or Specifications will be changed without noticedue to the improvement of performance or functions.