TROUBLESHOOTING
CHECK THA
T
FAUL
T
ST
AGE 2 CHECK
Pump shows no
indication that it is
powered up.
Mains plug is plugged in and
power switched on.
The power switch on the
pump is switched on.
The fuse in the mains plug is
not blown.
The wall socket that the
pump is connected to is
working correctly
.
1.
2.
3.
4.
Connect the pump to the
nearest (working) mains outlet.
Replace the plug fuse with the
correct 5A fuse.
Tr
y a different device in the
mains outlet.
1.
2.
3.
IF PROBLEM PERSISTS
Contact Har
vest Healthcare technical
support.
Before calling:
Please ensure you have the serial
number and model of equipment.
Please record details of the results of
the recommended tests. (Notes pages
are provided at the back of this user
manual).