Eastern Energy 840063 Instruction Manual Download Page 5

 

- 5 - 

B.  PEAK VALUE 

Switch the 

RMS/PEAK SELECTOR 

(A4) to the "PEAK" position.   

Peak value = 1.414 RMS value.

 

 
C.  DATA HOLD   

1.  Press the 

HOLD BUTTON 

(A3) during measurement to freeze the 

displayed value. "D.H" is shown on the LCD. 

2.  Press the 

HOLD BUTTON 

(A3) a second time to resume 

measurement. 

 

D.  DATA RECORD (Displays the Maximum and Minimum readings.) 

1.  To start the Data Record function, press the 

RECORD BUTTON

 (A5) 

once.  The LCD display will indicate "REC."   

2.  Press the 

RECALL BUTTON

 (A6) and the "Max" symbol (along with 

the maximum value) will appear on the LCD display.   

3.  Press the 

RECALL BUTTON

 (A6) again and the "Min" symbol (along 

with the minimum value) will appear on the display.   

4.  To exit the memory record function, press and hold the 

RECORD 

BUTTON 

(A5) for approximately 2 seconds. The display will revert 

back to the current reading. 

 

E.  AUTOMATIC SHUT OFF  

This unit has a built-in automatic shut off function to prolong battery life. 
The meter will shut off automatically if no buttons are pressed within 10 
minutes.  To deactivate this feature, select the memory record function 
during measurement, by pressing the 

RECORD BUTTON 

(A5). 

 

F.  RS232 PC SERIAL INTERFACE 

The unit features an RS232 output via 3.5 mm Terminal.  The connector 
output is a 16 digit data stream which can be utilized by user's specific 
application.  An RS232 lead with the following connection will be required 
to link the instrument with the PC serial input. 

 
The 16 digit data stream will be displayed as follows:  

Meter (3.5 mm jack plug) 

PC (9W "D" Connector) 

Center Pin 

Pin 2 

Ground/Shield 

Pin 5 

  D15D14D13D12D11D10D9D8D7D6D5D4D3D2D1D0

 

www.eastern-energy.net

Email: [email protected]

Summary of Contents for 840063

Page 1: ...VIBRATION METER 840063 Instruction Manual www eastern energy net Email info ete co th...

Page 2: ...vibration is a useful guide to the balance alignment and gen eral condition of pumps motors and other machinery Poor balance mis alignment and looseness results in higher levels of vibration and indic...

Page 3: ...dustrial vibration Switch the ACC VEL SELECTOR A7 to the VEL position 4 Switch the RMS PEAK SELECTOR A4 to the RMS position 5 Press the POWER BUTTON A2 once to turn on the meter 1 DISPLAY 2 POWER BUTT...

Page 4: ...non magnetic material hold the VIBRATION SENSOR and touch the tip to the object above b If the surface material is magnetic connect the VIBRATION SENSOR to the MAGNETIC BASE left and touch it to the o...

Page 5: ...the RECORD BUTTON A5 for approximately 2 seconds The display will revert back to the current reading E AUTOMATIC SHUT OFF This unit has a built in automatic shut off function to prolong battery life T...

Page 6: ...e are 4 different machine groups with 4 classification ranges and DO End Word D1 to D4 Upper Display reading D1 LSD D4 MSD D5 to D8 All D9 Decimal Point DP position from right to the left 0 No DP 1 1...

Page 7: ...ill permissible 4 51 11 2 mm s Dangerous 11 2 mm s GROUP T Largest machines and turbo machines with special foundations Good 0 2 80 mm s Acceptable 2 81 7 10 mm s Still permissible 7 11 18 0 mm s Dang...

Page 8: ...computer circuit V WARRANTY Sper Scientific warrants this product against defects in materials and workmanship for a period of one 1 year from the date of purchase and agrees to repair or replace any...

Reviews: