ET5301/5302/5303/5304 Programmable DC Electronic Load User Manual 

Hangzhou Zhongchuang Electronics Co., Ltd. 

 

 

 

Before starting the Tran test, relevant parameters of Tran test should be set, including: Tran state, 

level A, level A pulse width, level B, level B pulse width and Tran test mode. The setting interface and the 

testing interface are shown in figure 2.5.1 and figure 2.5.2 respectively.

 

Operating instructions: 1. Press

Mode

  to enter the main interface of Tran test, and press

Menu

to enter the parameter setting interface of Tran test. 2. Select the action item by turning the knob. 3. Press

Enter

to switch set options for non-numeric parameters.

 

3. As for the setting of numeric parameters, 

please refer to 1.4 numeric parameter settings. 4. Press

Esc

to return to the testing interface. 6. In the 

testing interface, press

On/Off

to turn on or off the load. 

 

   

 

Figure 2.5.1 Tran Set interface                                            Figure 2.5.2 Tran Test interface 

2.6 List Test   

The list test function makes it easy to test the working conditions of the source to be measured under 

different loading status, which is conducive to the automated testing of production line. Preset the steps of 

the list test, and then edit the test steps and test parameters of the source to be measured into a list and 

complete a series of tests in sequence. Specific parameter settings include: number of steps, step mode, 

repeat switch, and the load mode, load size, delay time, compare switch, maximum value and minimum 

value of each step. 

The setting interface and testing interface are shown in figure 2.6.1 and 2.6.2 respectively, and the 

test outcome interface at the end of the test is shown in figure 2.6.3.

 

Operating instructions:

 1. Press

Mode

  to enter the main interface of list test, and press

Menu

to enter the list menu interface, as shown in figure 2.6. For interface operation, please refer to the 

operating instructions for list menu above the figure. 2. Enter the sub-menu interface of list test setting, as 

shown in figure 2.6.1. For interface operation, please refer to the operating instructions for the list test 

setting interface above the figure. 3. After the completion of parameter settings, press

Esc

to return to the 

list menu interface, and then press

Esc

to return to the main interface of list test.4. In the testing interface, 

press

On/Off

to turn on or off the load. 5. After the completion of the electronic load startup test, the list 

Summary of Contents for ET5300

Page 1: ...ET5300 5301 5302 5303 Programmable DC Electronic Load User Manual Hangzhou Zhongchuang Electronics Co Ltd...

Page 2: ...Hangzhou Zhongchuang Electronics Co Ltd Adress NO 3 Kangle Road Gongshu District Hangzhou China Postcode 310015 Telephone 0086 571 56861333 Fax 0086 571 56861355 Website http www east tester com...

Page 3: ...ower output within a certain range Support 4 wire measurement The list test mode can simulate a variety of loading status changes The short circuit test can be used to simulate load short circuit Supp...

Page 4: ...tant power mode 6 2 4 5 Constant current to constant voltage mode 6 2 4 6 Constant resistance to constant voltage mode 6 2 5 Tran Test 7 2 6 List Test 8 2 7 Scan Test 11 2 8 Battery Test 12 2 9 LED Te...

Page 5: ...e 10000 S Li m i t 15 000 V 3 000 A 220 00 W V A W 10 00 A Icon bar there are five types of icons on the electronic load interface 1 2 Front panel key Range V 1 2 3 Mode 4 5 6 Enter 7 8 9 Local 0 ESC...

Page 6: ...s cannot be used Local For switching to the local mode Trigger Trigger key for manual trigger Keys used for moving the cursor left or right Range I Shortcut key for current range switching Range V Sho...

Page 7: ...n remote operation mode the corresponding icon will be displayed on the top of the interface At this time the interface is locked and the status and operation of the instrument can be controlled throu...

Page 8: ...CC CV constant resistance to constant voltage mode CC CR The parameter settings for the above six modes can be done in the measurement interface In addition press Mode on the main interface of any mod...

Page 9: ...r details 3 Press the corresponding On Off to turn on the load M EAS D I SP C V 0 000 0 000 Set 16 000 O f f Ti m e 10000 S Li m i t 15 000 V 3 000 A 220 00 W V A W 0 00 V Figure 2 4 2 Constant voltag...

Page 10: ...current to constant voltage mode is to prevent the source to be measured from being damaged by overcurrent discharge In this mode when the source to be measured cannot output the current value set by...

Page 11: ...rrents or voltages set by the load This function can be used to test the dynamic characteristics of the power source Moreover the Tran test is divided into three modes continuous mode pulse mode and t...

Page 12: ...mated testing of production line Preset the steps of the list test and then edit the test steps and test parameters of the source to be measured into a list and complete a series of tests in sequence...

Page 13: ...ion key to select save and then press Enter to enter the list test parameters saving interface as shown in figure 2 6 4 Please refer to the operating instructions for list test parameters saving inter...

Page 14: ...rface 1 Select the file by turning the knob 2 Press direction keys to switch to the edit state before any file can be edited Press the direction key to select the previous or next page and then press...

Page 15: ...igure 2 6 7 File rename interface 2 7 Scan Test Scan test can be used to detect the continuous working conditions of the source to be measured within a certain range and makes it easy to capture vario...

Page 16: ...to measure the discharge capacity of the battery Operating instructions 1 Press Menu in the main interface of battery test to enter the battery parameters setting interface 2 Select the action item b...

Page 17: ...ng the knob 2 For parameter setting please refer to 1 4 Numeric Parameters Setting for details 3 Press On Off to turn on or off the mode Option Description Name content Description LED Vo 0 Vmax Stead...

Page 18: ...rature protection and reverse polarity prompt the prompt information is subject to the real picture M EAS D I SP C C 0 000 0 000 Set 3 000 O f f Ti m e 10000 S Li m i t 15 000 V 3 000 A 220 00 W V A W...

Page 19: ...rted 2 15 Other system settings 2 15 1 Key Lock Function The load also added the lock function to prevent user misoperations The title bar will display the lock identifier In the lock out state all th...

Page 20: ...eness F1 The list test result is unqualified low effectiveness G GND Expansion port input N1 External trigger input The extension port output is OC gate output The user needs to connect the pull up re...

Page 21: ...ange 0 400W Resolution 10mW Accuracy 0 1 0 5 FS Tran Test Mode CC CV T1 T2 50ms 60s Battery Test Accuracy CC CR Discharge mode 9999Ah Maximum discharge capacity 1mA 10mA 10m 100m Range of measurement...

Page 22: ...Resolution 10mW Accuracy 0 1 0 5 FS Tran Test Mode CC CV T1 T2 50ms 60s Battery Test Accuracy CC CR Discharge mode 9999Ah Maximum discharge capacity 1mA 10mA 10m 100m Range of measurement Voltage read...

Reviews: