Environmental conditions for transportation and storage
Room temperature range for transportation
and storage
+4ºC to +40ºC
Relative humidity range for transportation
and storage
< 85% RH
Atmospheric pressure band
797 hPa to 1013 hPa
(598 mmHg to
760 mmHg)
Conditions for installation and operation of the equipment
Operating room temperature band
+10ºC to +30ºC
Operating relative humidity range (non-
condensing)
< 75% RH
Atmospheric pressure band
797 hPa to 1013 hPa
(598 mmHg to
760 mmHg)
Operating altitude
≤ 2000 m