e-tarti L60 Technical Manual Download Page 14

L60 & L60Tr Technical Manual  - 

L60 & L60Tr Teknik Manuel

  v1.0.5 

 

 

 

 

14 

 

6.

 

Troubleshooting 

Sorun Giderme

 

 

Sensor based questions (Abnormal 

measurement results) 

When the screen result is not shown as expected, reasons; 

 

No screen change on the display when loading or onloading 

the sensor: 

1.

 

Please check mV/V output of the sensor by using 

SERVICE->LC.TEST menu. Regular Loadcell 

sensors are uses 2mV/V and L60 device will show 

the measurement result as mV/V unit. So, there 
must be a visible change when loading and 

unloading conditions. If not please check that the 

electrical connections. Especially there must be a 

short circuit between “+Sense and +Excitation” 
terminals. And also a short circuit between “-

Sense and -Excitation” terminals must be done. 

Additionally, check that a break down on the 
loadcell cable all the way from sensor to L60 

device. Check that loadcell output is over loaded 

state (more than 3.9mV/V) or not too. If mV/V 

result on the LC.TEST menu is acceptable range 
(below +/-2.0mV/V) , then check that mechanical 

transitions from load platform to loadcell sensor. 

The loading must effect to the loadcell sensor, 
there can be an unwanted mechanical contact or 

comprension on the mechanical part 

2.

 

Check that the SENSOR-> GAIN is suitable with 

loadcell sensor (2mV/V)  

3.

 

If SERVICE->LC.TEST is success but there is no 

screen change, there is a calibration problem. 

Please make a calibration via CALIBRATION menu 
 

 

 

 

Err-01 (LC.MAXIMUM) or Err-05 messages 
Error-01 is related to devices Maximum parameter has been 

exceeded. So, please change DEVICE->MAXIMUM 

parameter. Or make a zero calibration, or press Zero button 
more than 1seconds 

Error-05 is related to loadcell sensor output has axceeded 

L60 devices measurement area (like L60 is configured to 
measure 2mV/V and the current input value is more than 

2mV/V). Please increase the mV/V selection on the Gain 

parameter (SENSOR->GAIN)  

 

 
 

 

Sensör tabanlı sorular (Anormal ölçüm) 

Ekranda gösterilen değer beklendiği gibi olmamasının nedenleri; 

 

Yük hücresine yükleme veya boşaltma yapılmasına rağmen 
ekran değişmiyor: 

1.

 

SERVİS->SENSÖR menüsüne giderek mV/V çıkış 
değerini kontrol ediniz. Normal Loadcell sensörleri 
2mV/V kullanır ve L60 cihazı ölçüm sonucunu mV/V 
birimi olarak gösterir. Bu nedenle yükleme ve boşaltma 
koşullarında gözle görülür bir değişiklik olmalıdır. 
Değilse, lütfen elektrik bağlantılarını kontrol edin. 
Özellikle “+Sense ve +Uyarı” terminalleri arasında kısa 
devre olmalıdır. Ayrıca “-Sense ve -Excitation” 
terminalleri arasında kısa devre yapılmalıdır. Ek olarak, 
sensörden L60 cihazına kadar yük hücresi kablosunda 
bir kopukluk olup olmadığını kontrol edin. Yük hücresi 
çıkışının aşırı yüklenmiş durumda (3,9mV/V'den fazla) 
olup olmadığını kontrol edin. LC.TEST menüsündeki 
mV/V sonucu kabul edilebilir aralıktaysa (+/-2.0mV/V 
'un altında) yük platformundan yük hücresi sensörüne 
mekanik geçişleri kontrol edin. Yükleme, loadcell 
sensörüne etki etmelidir, mekanik parça üzerinde 
istenmeyen bir mekanik temas veya sıkışma olabilir. 

2.

 

SENSOR->KAZANC  menusundeki kazancın 
kullanılan sensor ile uyumlu seçilip seçilmediğini 
kontrol edin (2mV/V)  

3.

 

Eğer SERVIS->YH.TEST menusundeki değerler 
uygunsa fakat ana ekrana dönüldüğünde değişim 
olmuyor ise kalibrasyon hatası 
vardır.KALİBRASYON menusunu kullanarak 
kalibrasyon yapınız 

 

Hata-01 (MAKSİMUM) yada Hata-05 mesajları 
Hata-01 maksimum parametresinin geçildiğini ifade eder. 

CİHAZ-MAKSIMUM parametresini artırınız. Eğer platform boş 

ise Sıfır Kalibrasyonu yapınız veya Sıfır tuşuna uzun basarak 
sıfırlama yapınız. 

Hata-05 Yük hücresi sensor girişinin aşırı değerde (ölçülebilir 

aalık dışında) olduğunu ifade eder. SENSOR->KAZANC 
parametresindeki mV/V değerini artırınız

 

 

 

Electrical Based Disturbances 

Power supply based issues are discussed in this topic. 

 

Please use a proper power supply which is 
sufficient enough to suppress electrical stress on 

power line and instantenous current loadings on 

the supply line 

 

If there is a pinomatic valves on the system and 

also auxiliary relays, there must be reverse 
diodes between the A1-A2 terminals (as short as 

possible to terminals) and the reverse diodes 

are also has to be places for pinomatic coil 

 Elektriksel Sorunlar 

Elektriksel bağlantı kaynaklı sorunlar bu başlık altındadır. 

 

Besleme hattında oluşabilecek anlık yükleme ve 
güç tüketimlerini karşılayabilecek bir güç kaynağı 
seçiniz 

 

Sistemde yardımcı role yada pinomatik valfler 
var ise A1-A2 terminalleri arasında ters diyot 
konmalıdır (klemens terminallerine müm 
mertebe en kısa şekilde olmalıdır). Pinomatik 
bobinleri uçlarına da aynı şekilde ters diyotlar 

Summary of Contents for L60

Page 1: ...L60 L60Tr Technical Manual L60 L60Tr Teknik Manuel v1 0 5 1 www etarti com e tart Technical Manual Teknik Manuel L60 17 03 2023 Last Updated...

Page 2: ...L60 L60Tr Technical Manual L60 L60Tr Teknik Manuel v1 0 5 2...

Page 3: ...esses Tart Program Adresleri 6 Device Commands Cihaz Komutlar 8 EEprom map 10 3 HOLD PROGRAM 11 Peak Hold Application ekme Basma Uygulamas 11 Inputs Giri ler 11 Outputs k lar 11 Hold Modbus Registers...

Page 4: ...digit LED Loadcell Y k H cresi 350 Ohm Maximum 16 Loadcell 350Ohm Maksimum 16 Y kH cresi Internal Resolution z n rl k 24bits 24Bit Measurement Speed l m H z 1280 samples second 1280 rnekleme saniye L...

Page 5: ...Gerilimi Supply Current Besleme Ak m Loadcell Y k H cresi 24V 30mA 1K 24V 50mA 45 Ohm 8x 350R loadcell 24V 100mA 16 Ohm 23x 350R loadcell 24V 150mA 11 Ohm 35x 350R loadcell 24V 200mA 8 2 Ohm 46x 350R...

Page 6: ...4 5 NA 6 Relay2 LED Relay is energized R le2 LED R le Enerjili 7 Relay1 LED Relay is energized R le1 LED R le Enerjili 8 10 3bit Decimal Point Noktan n yeri 11 Device language 0 English 1 Turkish Dil...

Page 7: ...Kullan m d RW 63 N A N A RW 64 75 N A N A RW 76 DAC Current Value DAC Anl k De er DAC Current Value Only DAC Filling 0 4095 Setting bigger values returns with 4094 to notify error DAC Anl k De er Sade...

Page 8: ...edilir Sadece Dolum parametreleri i in Hedef Hassas HedefDebi S f r Band Ba laGecikmesi Bo altmaBekleme Bo altmaGecikme DolumZamanA m Tolerans ve DaraLimit de erleri i indir DAC dolum i in de ge erlid...

Page 9: ...his cmd if not mandatory Eeprom yazma i lemi RegA Adres ve RegB Data yaz ld ktan sonra Komut adresine 2 yaz l r 3 Address Data Eeprom read operation Put RegA Address to be read and write the Cmd regis...

Page 10: ...o High order 10000d 2710h 56 16 57 39 58 59 0 Eeprom read write operations are 1byte on each write actions 68 Decimal Point Devices decimal point location 116 Baudrate 0 115200 1 57600 2 38400 3 19200...

Page 11: ...r R W Y O Address Adres Data Data Description Tan m RW 0 Command Komut Command Register Komut Yazmac RW 1 RegA Extension for command Komut i in yard mc yazma RW 2 RegB Extension for command Komut i in...

Page 12: ...Enabled Uzama Aktif 10 15 N A reserved RW 6 7 Distance Encoder Count Mesafe Sayac Number of counts from input2 to calculate moving distance Giri 2 de g zlenen darbe adedi sayac d r Mesafe izleme ama l...

Page 13: ...000 counts As LED screen The examples below are given for 2mV V input Coefficient Screen 1 00000 430000 0 10000 43000 0 01000 4300 0 02326 10000 0 00698 3000 5 Error Messages Codes Hata Mesajlar Kodl...

Page 14: ...n r ve L60 cihaz l m sonucunu mV V birimi olarak g sterir Bu nedenle y kleme ve bo altma ko ullar nda g zle g r l r bir de i iklik olmal d r De ilse l tfen elektrik ba lant lar n kontrol edin zellikle...

Page 15: ...cient enough to supply more then twice power in Watt konmal d r Bobinler enerjilendiklerinde veya enerjileri kesildiklerinde 2kV a kadar anl k gerelim retirler ve bu gerilimler sisemi etkilerler A ma...

Reviews: