L60 & L60Tr Technical Manual -
L60 & L60Tr Teknik Manuel
v1.0.5
13
4.
Coefficient Calibration
Katsayı Kalibrasyonu
On the LED screen the internal counts are shown as 1/10 with scaling 0 to 430,000 counts for 2mV/V
The coefficient is same as screen value and 0.1 is cause to be a calibration to 43.000 screen value for 2mV/V
0mV/V --> 0 counts
2mV/V--> 430.000 counts (As LED screen)
The examples below are given for 2mV/V input
Coefficient
Screen
1.00000
430000
0.10000
43000
0.01000
4300
0.02326
10000
0.00698
3000
5.
Error Messages-Codes
Hata Mesajları-Kodları
Error No
Hata No
Description
Tanım
Meaning
Anlamı
ERR.005
"LC.HATA"
Loadcell is overloaded or not installed. Check the connection and the
load value
ERR.090
"ADCHat "
ADC unit of the device is broken. Needs service
ERR.001
"MAK.YUK"
The screen value is above the Maximum Capacity
ERR.099
"DEG.COK"
Parameteter Error, entered value is above the limits