- 3 -
The e-medic Silence TP 2 & TP 1are intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The user of the
e-medic Silence TP 2 & TP 1 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the e-medic Silence TP 2 & TP 1
as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of
transmitter
m
150 kHz to 80
MHz
d =
1,2
80 MHz to
800 MHz
d =
1,2
800 MHz to
2,5 GHz
d =
2,3
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the
recommended separation distanced
in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where
P
is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Guidance and manufacturer’s declaration – electromagnetic immunity
The e-medic Silence TP 2 & TP 1 are intended for use in the electromagnetic
environment specified below. The user of the e-medic Silence TP 2 & TP 1
should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment
–
guidance
Electrostatic
Floors should be wood,