    Measuring Transducer TK-101     

Mounting- and Operating Instructions

Ex-Zone 0 Category 1

Important safety instructions please read and note

If the transducer is to operate smoothly and safely,  it is vital that it is transported,  stored and
assembled appropriately, installed and commissioned correctly, used only as intended and repaired as

required.   These   actions   should   only   be   performed   by   individuals   with   the   required   specialist
knowledge and corresponding qualifications.
The applicable safety regulations for the erection and operation of electrical systems in explosion risk

areas shall be observed. Your particular attention is drawn to the erection provisions in accordance
with 

EN 60079-0, EN 60079-11, EN 60079-26, 

for electrical plants in areas subject to explosion

hazards.   The   attached   EC-type   examination   certificate  

TÜV   02   ATEX   1795   X  

respectively

IECEx TUN 17.0036X

 must also be observed.

Should the information contained in the instructions below prove to be inadequate in any way, please

contact the manufacturer.

Application

The TK-101 transducer is integrated directly into the connection socket of our continuous filling level

probes of the type series TK-30... It converts the resistance signal of the continuous level measuring
sensors into a unit current signal (4..20 mA) proportional to the filling level height. installed in the

float. The TK-101 has a potentiometer for zero adjustment (MIN), an for 100% setting (MAX).

Mounting

The TK-101 is intended for the installation into a housing, which offers suitable protection against the
respective  ambient conditions  during  operation.  At  the factory,  the TK-101  is integrated into  the

respective TK-30_ housing.

Electrical connection 

If the probe is used in hazardous areas, it may only be operated via a Zener barrier or an approved
Ex-supply device. The electrical connection must be made without power.

Version 05/03/2018

Errors and changes reserved