7/08/14
10:00
Measurements
Ambient CO
CO
ppm
0
CO max
ppm
0
It is mandatory to perform the
instrument autozero in fresh, clean
outdoor air.
7/08/14
10:00
Measurements
Ambient CO
CO
ppm
0
CO max
ppm
0
Connect the gas probe to the instrument and perform
the measurement.
Wait 5 minutes.
7/08/14
10:00
Measurements
Draft
P diff
inH2O
0.000
O
7/08/14
10:00
Measurements
Draft
P diff
inH2O
- 0.020
O
Once the
pressure zeroing
is completed
insert the probe in
the chimney and
measure the draft.
AA SF--A
AAC TA03A
AAC KP01
AA SA08
7/08/14
10:00
Measurements
Draft
P diff
inH2O
0.001
O
Before zeroing the
pressure it is
mandatory to
REMOVE the
probe from the
chimney.
View & Save the
DRAFT
measurement.
ZERO Draft
Sensor
O
Save the
measurement
Draft
Ambient CO
7/08/14
10:00
Eff. tot
%
91.4
Loss tot
%
8.6
ΔT
°F
74.7
T air
°F
15.4
T flue
°F
190.1
Ex. Air
1.25
CO
2
%
9.3
O
2
%
4.2
Combustion analysis
UNI 10389
KEEP
PRESSED
FOR 2
SECONDS
UNTIL IT
BEEPS
WARNING!
During autozero DO NOT
insert the gas probe in the
chimney.
WARNING!
Be sure all connections are tight to assure accurate sampling.
It is important that water trap & filter be installed vertically to
prevent damage from moisture & particles to sensors.
When conducting measurements, the water trap/filter
assembly MUST be in a VERTICAL position.
When testing is completed, always drain the water
trap with any condensation (after EVERY test!).
AA AL05
Configuration
Air temperature
15/01/1
4
Probe T
°C
25.0
Air T
°C
-----
Air temp.
Analysis
While in manual analysis, holding pressed both keys and
makes the instrument switch off the suction fumes
pump and blocks the refresh of any current measure.
To switch on the suction fumes pump again and reactivate
the refresh of the current measure, press again the keys
and .
Configuration
Air temperature
15/01/1
4
Probe T
°C
25.0
Air T
°C
25.0
It saves the value,
acquired or entered
in
the
parameter
‘Air T’.
Configuration
Air temperature
15/01/1
4
Probe T
°C
25.0
Air T
°C
-----
A c q u i r e s
t h e
temperature
value
detected from the
sampling probe. That
value is reported in
the parameter ‘Air T’.