24
6.2.5
Fill with distilled water.
Fill the humidity calibrator with max. 1300 ml distilled (deionized) water.
If the max. fill lever is exceeded, Water - Level high is displayed on the
display and water must be removed until the message disappears.
(see Item 8.2)
During filling, pay special attention that no water enters into the
measuring chamber.
6.2.4
Open water container.
Open screw plug with the help of the included Allen key.
6.2.6
Close water container.
Close screw plug with the help of the included allen key.
6.2.7
Wait through warm-up phase.
Observe 20 min warm-up phase (Display: WARM UP TIME).
6.2.9
HUMOR is ready for operation.
6.2.8
Establish media supply (oil-free compressed air or N2).
Establish media supply and charge humidity - generator with pressure.
(p < 10 bar
(145psi)
see safety tips)
english