26
User Manual EE741 Inline Flow Sensor for Compressed Air and Gases
10.1 Factory Settings of the Outputs for DN15 / DN20 / DN25
Analogue output
Switch output
hysteresis mode
Minimum flow
shutdown
Pipe
diameter
from
to
SP
HY
SP
HY
Standard volume flow [Nm³/h]
DN15
0
75
50
5
0.15
0.07
DN20
0
130
90
9
0.25
0.12
DN25
0
200
150
15
0.35
0.17
Standard volume flow [Nm³/min]
DN15
0
1.25
0.83
0.08
DN20
0
2.15
1.5
0.15
DN25
0
3.3
2.5
0.25
Standard volume flow [l/min]
DN15
0
1250
833
83
DN20
0
2150
1500
150
DN25
0
3300
2500
250
Standard volume flow [l/s]
DN15
0
20
14
1.4
DN20
0
35
25
2.5
DN25
0
55
40
4
Standard volume flow [SCFM]
DN15
0
44
30
3
0.15
0.07
DN20
0
76
53
5.3
0.25
0.12
DN25
0
117
88
8.8
0.35
0.17
Mass flow [kg/h]
DN15
0
97
65
6.5
DN20
0
165
115
11.5
DN25
0
255
195
19.5
Mass flow [kg/min]
DN15
0
1.6
1
0.1
DN20
0
2.8
2
0.2
DN25
0
4.3
3.2
0.32
Standard flow [Nm/s]
DN15
0
120
80
8
DN20
0
120
80
8
DN25
0
120
80
8
Standard flow [SCFM]
DN15
0
23600
15000
1500
DN20
0
23600
15000
1500
DN25
0
23600
15000
1500
Temperature [°C]
DN15
-20
60
24
0.5
DN20
-20
60
24
0.5
DN25
-20
60
24
0.5
Temperature [°F]
DN15
-4
140
75
1
DN20
-4
140
75
1
DN25
-4
140
75
1
Pulse output:
Pulse duration:
0.1 s
Pulse value:
1 m³
Measured value averaging:
10