Manual english
EE381
page 26
Continuing self-diagnostic:
Definition:
•
Error:
- possible cause
=> measures / help
•
Display shows incorrect values:
- Error during re-adjustment of the transmitter
=> Reset to factory calibration and repeat the
calibration routine
- Output configured incorrectly
=> Check output range and output signals
in the configuration
•
Transmitter failure:
- No supply voltage
=> Check wiring and supply voltage
=> Only green LED is illuminated continuously => electronics
defect => contact the manufacturer