User Manual EE360
Moisture in Oil Sensor | 27
6.2 3.5" TFT Colour Display (optional)
The EE360 display includes a data logger and push buttons for full configuration of the device. Upon start-up of
an EE360 with display, the data logger and the configuration menu will be initialized during the first 5 seconds.
Chart + data logger
Configuration menu
Status information
Buzzer ON / OFF
Error indication / information
Display settings
Fig. 23 Display with push buttons
6.3 Chart and Data Logger
The TFT display with the integrated data logging function saves all measured and calculated values to the internal
memory. The data logger has a real time clock (UTC time) with a battery back-up.
NOTICE
Changing the UTC time erases all stored data.
The data logger can save 20 000 values for each measurand. The logging interval is user configurable from 1
second to 12 hours. The data logger menu is also used to select the data points that make up the graph and for
scaling, see chapter 10 Appendix, Fig. 24 Data logging).
Home button
Configuration incl. data logger menu
Time scale for the graph (5 min - 3 months)
Cursor
Axis 1 value
(cursor position)
Axis 1
Axis 2
Axis 2 value
(cursor position)
Date
(cursor position)
Time
(cursor position)
Fig. 24 Data logging
Each point in the graph represents a logged value. The points are connected by a linear interpolation.