36
Configurations software
5.4
Calibration
In addition to the manual calibration procedure on the circuit board (see
Hardware,
chapter 7 “Humidity/Temperature Calibration”), new calibrations can be
performed using the EE31 software.
Note: A two-point calibration for temperature is only possible on the circuit board
and is not supported by the software.
1-point calibration
Humidity
: Fast and easy calibration for accurate measurement results at a defined working
point (humidity point).
For calibration procedure see Hardware, chapter 7. “Humidity/Temperature Calibration”
1) Stabilise the probe of the desired humidity for min. 30 minutes.
2) Click on the Humidity "1-point calibration" button. The measured values will
now appear in both input fields.
3) Replace the value in the input field "Humidity Reading" with the reference
humidity (value of the saline solution or display of HUMOR 20).
4) By clicking on "Save", the humidity reading for the transmitter will be
adjusted to the reference humidity.
5.3
Sensor / Probe Replacement
In case of sensor or probe replacement, a costumer calibration has to be done.
Replacement -
Humidity Sensor (RH):
1) Replace the humidity sensor by a new one
(
see Hardware, chapter 8.1 Sensor Replacement)
.
2) Run the humidity calibration
(see chapter 5.4)
.
Replacement - Probe (RH & T): 1) Replace the humidity sensor by a new one
(see Hardware, chapter 8.2 Probe Replacement.
2) Run the humidity and temperature calibration
(see chapter 5.4)