8
User Manual EE300Ex Humidity / Temperature Sensor
2.3 Certification
EUROPE:
The EE300Ex sensor fulfills the
ATEX Directives
on intrinsically safe operating equipment.
Applied standards for ATEX:
•
EN 1127-1:2011
•
EN 60079-0:2012
•
EN 60079-11:2012
The EU-Type Examination has been carried out by TÜV SÜD Product Service GmbH.
Certified to EU-Type Examination
TPS 13 ATEX 38892 003 X
.
Entity parameters U
i
= 28V; I
i
= 100mA; P
i
: = 700mW; C
i
= 2,2nF; L
i
≈ 0mH
Ex-Designation
Sensor without display
I II 1G Ex ia IIC T4 Ga / II 1D Ex ia IIIC T80°C Da
Sensor with display
I II 2G Ex ia IIC T4 Gb / II 1G Ex ia IIB T4 Ga
Remote probe
II 1G Ex ia IIC T6-T1 Ga / II 1D Ex ia IIIC T80°C...200°C Da
Working temperature range for the probes:
Specification of the temperature class „TKG“ for use in gas area exposed to explosion hazards and
„TKD“ for use in dust area exposed to explosion hazards as a function of the ambient temperature
„Tamb“ for the humidity and temperature probe and the temperature probe:
INTERNATIONAL:
Applied Standard for IECEx:
•
IEC 60079-0:2011
•
IEC 60079-11:2011
The Certificate of Conformity has been carried out by FM Approvals.
Certificate No.:
IECEx FMG 14.0017 X
Entity parameters: 6.4 Vdc ≤ U
i
≤ 28Vdc; I
i
= 100mA; P
i
= 700mW; C
i
= 2,2nF; L
i
= 0mH
Ex-Designation
Sensor without display
Ex ia IIC T4 Ga / Ex ia IIIC T131°C Da
Sensor with display
Ex ia IIC T4 Gb / Ex ia IIB T4 Ga
Remote probe
Ex ia IIC T6-T1 Ga / Ex ia IIIC T80°C Da
Humidity and temperature probe:
•
T6 temperature class based on -40°C (-40°F) ≤ Ta ≤ 60°C (140°F)
•
T5 temperature class based on -40°C (-40°F) ≤ Ta ≤ 75°C (167°F)
•
T4 temperature class based on -40°C (-40°F) ≤ Ta ≤ 110°C (230°F)
•
T3 temperature class based on -40°C (-40°F) ≤ Ta ≤ 175°C (347°F)
•
T2 temperature class based on -40°C (-40°F) ≤ Ta ≤ 180°C (356°F)
• T1 temperature class based
on -40°C (-40°F) ≤ Ta ≤ 180°C (356°F)
TKG
TKD
Humidity and
Temperature Probe
TKG
TKD
Temperature Probe
T6
80°C
-40°C ≤
Tamb
≤ +60°C
T6
80°C
-70°C ≤
Tamb
≤ +60°C
T5
95°C
-40°C ≤
Tamb
≤ +75°C
T5
95°C
-70°C ≤
Tamb
≤ +75°C
T4
130°C
-40°C ≤
Tamb
≤ +110°C
T4
130°C
-70°C ≤
Tamb
≤ +110°C
T3
195°C
-40°C ≤
Tamb
≤ +175°C
T3
195°C
-70°C ≤
Tamb
≤ +175°C
T2
200°C
-40°C ≤
Tamb
≤ +180°C
T2
220°C
-70°C ≤
Tamb
≤ +200°C
T1
200°C
-40°C ≤
Tamb
≤ +180°C
T1
220°C
-70°C ≤
Tamb
≤ +200°C
Summary of Contents for EE300Ex
Page 31: ...31 User Manual EE300Ex Humidity Temperature Sensor 8 ATEX Certificate...
Page 32: ...32 User Manual EE300Ex Humidity Temperature Sensor...
Page 33: ...33 User Manual EE300Ex Humidity Temperature Sensor...
Page 34: ...34 User Manual EE300Ex Humidity Temperature Sensor...
Page 35: ...35 User Manual EE300Ex Humidity Temperature Sensor 9 EU Declaration of Conformity...
Page 51: ...51 User Manual EE300Ex Humidity Temperature Sensor 13 Control Drawing M1_1309080...
Page 52: ...52 User Manual EE300Ex Humidity Temperature Sensor...
Page 53: ...53 User Manual EE300Ex Humidity Temperature Sensor...
Page 55: ...55 User Manual EE300Ex Humidity Temperature Sensor...