23
User Manual EE300Ex Humidity / Temperature Sensor
4.3 Grounding and Potential Equalization
The EE300Ex shall be integrated into the potential equalization to avoid hazards from electrostatic
charges. The grounding shall comply with EN60079-14, EN60079-25 or IEC60079-14, IEC60079-25. A
remote probe shall also be grounded using a screw connection with a maximum of 1 MΩ in the potential
equalization.
The ground conductor or the potential equalization connection must have a cross-section of 4 mm²
(0.06 in²) for the external grounding. Stranded wires shall be fitted with suitable wire-end sleeves.
The resistance of the ground connection of the intrinsically safe barrier shall be max. 1 Ohm.
For installation in the U.S. use internal ground connection. The wire cross-sectional must comply with
NEC Section 250.122.
Cable cross-section
maximum 4mm² (0.06in²)
Cable cross-section
maximum 10 mm²
(0.2 in
2
)
Grounding connection channel 1
Grounding connection channel 2
Grounding connection occupied
External grounding:
Internal grounding:
Summary of Contents for EE300Ex
Page 31: ...31 User Manual EE300Ex Humidity Temperature Sensor 8 ATEX Certificate...
Page 32: ...32 User Manual EE300Ex Humidity Temperature Sensor...
Page 33: ...33 User Manual EE300Ex Humidity Temperature Sensor...
Page 34: ...34 User Manual EE300Ex Humidity Temperature Sensor...
Page 35: ...35 User Manual EE300Ex Humidity Temperature Sensor 9 EU Declaration of Conformity...
Page 51: ...51 User Manual EE300Ex Humidity Temperature Sensor 13 Control Drawing M1_1309080...
Page 52: ...52 User Manual EE300Ex Humidity Temperature Sensor...
Page 53: ...53 User Manual EE300Ex Humidity Temperature Sensor...
Page 55: ...55 User Manual EE300Ex Humidity Temperature Sensor...