22
12.3 Outputs
Each analogue output can be configured by clicking on “Edit”:
In general there are two methods to map a measurement signal to an analogue output:
(a) “mapping the special port ‘X’ of a transmitter ‘Y’ to an analogue output”.
This configuration always maps the measurement signal of port ‘X’ to the output. In addition, it does not matter if
the sensing probe is replaced by another one of the same type (e.g. calibration cycle or replacement of the probe).
(b) “mapping a specific sensing probe (with a defined serial number) to an analogue output.
This configuration maps ONLY this specific sensing probe to the analogue output. In addition, it does not matter
to which transmitter the sensing probe is a connected.
1. Mapping the selected measurement signal:
Example (a): “mapping an EE07 sensing probe at port 2 of the transmitter ‘10045689788’
Assigned to:
Select “Transmitter”.
Serial Number:
Enter the serial number of the desired transmitter.
The serial number can be copied and paste from the transmitter list (see chapter 12.2.1
Transmitter List) with ‘Ctrl’+’C’ and ‘Ctrl’+’V’.
Transmitter Port: Select the transmitter port (see chapter 12.2.2 Transmitter Details, under “TM-Port”) to which the
sensing probe is connected.
(Port 1, 2 or 3, resp.” 255”, if only one port is occupied, but the port number is unknown)
Example (b): “to map an EE07 sensing probe ‘0909500001055D’
Assigned to:
Select “Probe”.
Serial Number
Enter the serial number of the sensing probe.
The serial number can be copied and paste from the “Transmitter Details” of the transmitter list
(see chapter 12.2.1 Transmitter List) with ‘Ctrl’+’C’ and ‘Ctrl’+’V’.
Transmitter Port: Enter “255”.
2. Setup of ranges and values:
Measurand and Unit: Selected desired measuring value (T, RH …).
Physical Range: Enter desired range and engineering unit of the measured value (e.g. 0 ...100 °C).
Output Type:
Indicates the physical output quantity (factory setting according to order code).
Custom Range: Analogue output value, to represent the “Physical Range” as indicated above.
Failsafe Value:
Default analogue output value if a transmission failure / alarm is present.
1. Mapping the selected measurement signal
2. Setup of ranges and values