52
Rev. 0.7
[email protected] • www.e-breathe.de
53
Rev. 0.7
e-breathe e-Flow
[email protected] • www.e-breathe.de
e-breathe e-Flow
5.10 Charging Station
Display Symbol
Meaning
• red LED - permanent lighting
= No battery is inserted in the charging station.
= A battery is inserted, but it is not being charged due to the wrong voltage.
• green LED - permanent flashing
= Battery is charging
-> Flashing at short intervals = Battery capacity at 0- 33
-> Flashing at medium intervals = Battery capacity at 33 - 66
-> Flashing at long intervals = Battery capacity at 66-99%
• green LED - permanent lighting
= Battery capacity completely at 100%.
Meaning
Cause of error
Action
Display Icons
red LED -
2x continuous flashing
The battery is inserted
but the power supply of
the charging station is not
plugged in.
Remove the battery from
the charging station. Con-
nect the charging station
to the AC adapter or power
source.
red LED -
3x continuous flashing
Battery error code: critical
temperature is reached
during charging.
Remove the battery from
the charging station and
have it inspected by the
manufacturer.
red LED
4x permanent flashing
Input voltage too low to
charge.
Disconnect the power sup-
ply unit from the power
source and restart the char-
ging process. If the fault still
occurs, have it checked by
the manufacturer.
5.10.1 Charging Station Error Codes
The e-Flow charging station is required to charge the e-Flow battery.
The charger may only be used to charge the e-Flow battery for the e-Flow blower system. The e-Flow battery can only
be charged with the original e-Flow charging station from e-breathe.
The charging station is intended exclusively for indoor use in buildings and may only be used if the battery to be
charged is intact, the charging station is intact and the corresponding power supply unit is intact.
The charging station is equipped with safety electronics and software that monitors the charging process of the
battery and shows the user the state of charge of the battery. This is done via a coloured LED on the charging station.
Furthermore, it indicates via the different error codes if the battery is incorrectly charged and/or the electronics of the
battery is defective.
6.0 Maintenance and Cleaning
The plan specifies the minimum requirements for maintenance routines to ensure that you always have proper wor-
king equipment.
Maintenance and cleaning can only be carried out by appropriately trained personnel who are well-trained in
the nature of the task.
6.1 Maintenance
system
component
work to be performed
Breathing
conn
breathing air hose
See instructions for use of the breathing connection.
Filter
Expiration date check
x
Visual inspection
x
Filter capacity check
x
x
Blower Unit
(incl. battery and
charging station)
Visual inspection by the user
x
x
Battery level check
x
x
Battery charging
x
x
x
x
Battery replacement
x
Filter replacement
x
Air flow and warning devices check
x
Replacement of seals
x
x
Cleaning and disinfection
x
x
x
Service and maintenance performed by e-breathe Service
x
Bef
or
e use
A
ft
er use
A
nnually
Ev
er
y 2 y
ears
If nec
essar
y
To ensure the functionality of the blower, the unit must be serviced annually by an authorised service partner or by
the manufacturer. Authorized service partners can be found at: www.e-breathe.de
The next service date can be found in the system information menu of your blower.
Only use original e-breathe products. Do not make any changes to the equipment. The use of non-original parts
or modifications to the blower may reduce its protective function, compromise product approvals, and perman-
ently damage the blower unit. Non-compliance will void the warranty.