11. Measurement of SS
Caution
■
Be sure to perform calibration before installation
-
When power is initially supplied, SS density values will appear real time on the FND.
-
When the RUN lamp blinks, it indicates that analog output (4~20mA) is generating
-
When it does not blink, the analog signal is not being produced.
-
At the time of entry into MODE function, the analog output (4~20mA) will be put on hold
automatically
12. Measurement of Temperature
1) Compensation of Automatic Temperature Compensation (ATC)
-
This function should be carried out when the calibration of temperature is needed.
Pressing the key three times will make C-1 Mode appear
The figures will blink when the key is pressed
You can change the temperature by selecting either increase or decrease in temperature and pressing
the key one at a time slowly
When completed, please press the key
2) Input of Manual Temperature Compensation (MTC)
-
This function should be carried out when the calibration of temperature is needed.
The C-2 Mode will be displayed when the key is pressed four times
The temperature value will blink when the key is pressed.
Input a value using and keys.
When completing the input of data, press the key
※
While the temperature is input on C-2 Mode, the ATC function will come to a stop.