M0726(2):LMFI Series 5000 Lab Melt Flow Indexer
15
English 02/2014
Level the Melt Indexer
Using a small round bubble level, level the melt indexer. Place the level on top of the COLD barrel
and using the adjustable screw feet, bring the machine into level. Tighten the locking nuts to keep
the feet in level position Test shake the melt indexer for stability.
Some companies may choose to bolt the machines directly to their benches.
This can be done by removing the levelling feet and using their mounting holes to mount the
instrument to a bench or table. Use caution and remove power before doing so. Mounting this
way will entail removing some covers for internal access to the internal, main instrument plate.
Be sure to remove the level before turning on the machine. The level will be damaged if it gets hot.