background image

7

Dynatron® ibox™ Operator’s Manual  

|

  Rev. 2  

|

  5/14/2019

The Dynatron® ibox™ Iontophoresis Device

table of contents

symbols & labels

limited warranty

prev  

<

>

  next

prev topic 

<

 

<

>>

  next topic

The Dynatron ibox power may be turned OFF before, during, or after 
a treatment.  Depressing the ON/OFF switch during a treatment 
immediately ramps down the delivered current intensity to 0.5 mA. 

2. 

DISPLAY WINDOW:

  The Dynatron ibox provides information to the 

user through its Organic Light Emitting Diode (OLED) display window.  
The display window will progress through a sequence of Polarity, 
Dose, and Intensity during the setup phase of the treatment.  During 
the treatment phase, all three parameters will appear in the display 
window.

Polarity: 

 The display window allows the user three options for 

polarity:  Positive (+), Negative (-), and Dual Polarity (+/-).  Positive 
(+) delivers a positive current to the active electrode during a single 
phase treatment.  Negative (-) delivers a negative current to the active 
electrode during a single phase treatment.  Dual Polarity (+/-) delivers 

a positive current to the active electrode first, followed by a negative 

current.
 

Dose: 

 The display window shows a treatment dose as either a 

single number during the setup stage or as a ratio of dose delivered 
compared to the target dose in the treatment stage.  The dose is 
displayed in milliampere-minutes (mA-min) and is followed by a “+” or 
“-” sign to indicate the polarity of the treatment.  Below is an example 
of a dose ratio displayed during a treatment.
 

6/40+

This figure indicates that 6 mA-min of dose has been delivered and 

that the total target dose for the treatment is 40 mA-min.  The (+) 
indicates that this treatment has a positive polarity.

Time:

 The remaining time for a treatment is displayed in standard 

time format using a colon to separate minutes from seconds as shown 
below.

14:33

Time always counts down.  Time is always calculated according to 
the amount of dose to be delivered and the intensity of the current.  
When a treatment is completed, time will be displayed as “00:00.”

Intensity:

  Intensity is the amount of current delivered to the 

electrodes during a treatment.  Intensity is displayed with two digits 
separated by a decimal point and followed by the polarity of the 
treatment as shown below.  The units used are milliamperes (mA).  
Current intensities can typically range from 0.5 to 4.0 mA.

3 .8 mA+

3. 

KEYPAD:

  User control of the Dynatron 

ibox is through interactions with the 
keys on the keypad.  All control is done 
with single depressions of individual 
keys.  Some of the keys may be held 
down to allow for successive increases 
or decreases of certain parameters such 
as dose and intensity without having 
to press and release the key for each 
change.

Note:  If the Dynatron

®

 ibox

 is not actively delivering a treatment and 

remains idle for 60 seconds, the device will automatically turn off.

Summary of Contents for Dynatronics ibox

Page 1: ...1 Dynatron ibox Operator s Manual Rev 2 5 14 2019 table of contents symbols labels limited warranty next prev topic next topic ...

Page 2: ... 5 14 2019 table of contents symbols labels limited warranty prev next prev topic next topic OPERATOR S MANUAL The Dynatron ibox Iontophoresis System is a non invasive drug delivery system designed as an alternative to hypodermic injections ...

Page 3: ...lectrical Shock Hazard Contact with other sources of electricity may pose a risk of electric shock Do not connect this deice to any source of electricity IMPORTANT Before treating a patient with the Dynatron ibox read the entire Operator s Manual paying specific attention to Contraindications Warnings and Precautions COMPLIANCE The contents of this Instructions For Use manual are exactly the same ...

Page 4: ...s 16 Quick Reference Single Phase Treatment Instructions 19 Quick Reference Dual Phase Treatment Instructions 20 Contraindications Precautions and Warnings 21 Contraindications 21 Warnings 22 Precautions 23 Adverse Effects 23 Technical Information 24 Outputs 24 Treatment Timing 24 Dimensions 24 Environmental Conditions 24 Electromagnetic Emissions and Immunity 25 Batteries 29 Maintenance and Repai...

Page 5: ...carrier and take any steps necessary to file a claim for the damage sustained If possible use the original shipping carton if the device must be shipped for any reason Improper packaging of the device during transport can result in damage and will invalidate the warranty Complete the warranty registration form and return it to Dynatronics within 30 days of purchase This is to insure that you are n...

Page 6: ...om left of the keypad and is shown by the international ON OFF power symbol Power ON Press and hold the symbol until a beep is heard and the display illuminates When first powered on the Dynatron ibox logo will be displayed followed by the software version number The initial screen displayed after power on allows the user to choose between a default treatment or a custom treatment Power OFF Press ...

Page 7: ... milliampere minutes mA min and is followed by a or sign to indicate the polarity of the treatment Below is an example of a dose ratio displayed during a treatment 6 40 This figure indicates that 6 mA min of dose has been delivered and that the total target dose for the treatment is 40 mA min The indicates that this treatment has a positive polarity Time The remaining time for a treatment is displ...

Page 8: ...n of this manual for additional information on the Dynatron ibox batteries 5 OUTPUT JACK The output jack allows for connection of the electrode cable The jack only accepts an authorized Dynatron ibox electrode cable The electrode cable plug is keyed so that it must be orientated properly before insertion into the jack 6 BATTERIES Two 1 5 Volt AA Alkaline batteries are used to power the Dynatron ib...

Page 9: ...user to be captured as part of a treatment setup After depressing of this key the parameter just modified will be captured and the next treatment parameter to be defined will be displayed After all parameters have been entered pressing the Up Arrow key starts the treatment An enter key is not used during a treatment However depression of an enter key when not delivering a treatment will cause the ...

Page 10: ...ut to the patient electrodes 12 ACTIVE ELECTRODE CLIP The ACTIVE clip end of the electrode cable is light gray The clip attaches to the active electrode on the patient and may be squeezed in order to attach to the electrode more easily or it can be pushed onto the electrode without squeezing 13 RETURN ELECTRODE CLIP The RETURN clip end of the electrode cable is dark gray in color The clip attaches...

Page 11: ...es anywhere along the path of the wires including the electrodes clips and plugs Solutions 1 Check that the electrodes are not touching each other on the patient 2 Check that the electrode clips are not touching each other 3 Carefully inspect the entire electrode cable for any shorted wires caused by breaks in the wire Replace the electrode cable if any break or damage is found The treatment conti...

Page 12: ...during a treatment the battery voltage is less than 2 1 Volts During display of this message eight long beeps will be generated by the internal speaker Following these beeps the Dynatron ibox will no longer be operational i e you may not setup a treatment or if a treatment is in progress it will be discontinued Solutions 1 If message occurs with power on replace both batteries and retry powering o...

Page 13: ... not shut down due to the over current condition MSP430 ERROR This error message is displayed if the internal mi crocontroller circuit encounters difficulty controlling the device If this condition occurs any treatments in progress will be discontinued and the outputs turned off Although this is a rare condition if it is recurrent the user should not use the device for any further treatments and D...

Page 14: ...annel output jack located at the top of the device The plug end of the cable is keyed and must be oriented appropriately before it can be inserted into the jack Do not force the plug into the jack 3 Prepare the electrode sites on the patient Use an alcohol preparation pad to gently remove any dirt and or oils at the site of the intended electrode placement on the patient Make sure that the area pr...

Page 15: ...edefined treatment or a custom treatment Default Treatments Default Treatments are commonly used treatment setups that have been previously stored by the user from prior treatments or manufac turer s settings Default treatments allow for rapid treatment setups without having to enter treatment polarities modes and doses These treatments are stored inside the device s microcontroller and are not af...

Page 16: ...pt this polarity mode 12 Enter a custom dose s For a single phase treatment use the up and down arrow keys to adjust the dose of the treatment to the desired amount The dose can be adjusted from 1 to 160 mA min Holding down an up or down arrow key accelerates the dose value up or down so that the user does not have to press a key for each increment of dose When the desired dose is finally chosen p...

Page 17: ...fifteen seconds At the completion of a single phase treatment the intensity current will be automatically ramped down by the device to 0 mA over a period of about ten seconds At this time the device will generate four beeps followed by two beeps to signal the end of the treatment Dual Phase The display shown on the right is a typical display of a dual phase treatment currently occurring and shows ...

Page 18: ...the memory of the device s microcontrol ler for use as Default Treatments in the future This allows the user to recall and start a commonly used treatment on a channel without having to enter the treatment parameters To store a channel s default treatment the user must wait for a treatment to complete The user then presses the up arrow key The display below will be seen by the user If the user pre...

Page 19: ...ctions table of contents symbols labels limited warranty prev next prev topic next topic Quick Reference Single Phase Treatment Instructions Select Dose UP DOWN ARROW KEYS ENTER During Treatment Select Polarity UP ARROW KEY ENTER Pre Launch UP ARROW KEY OR ENTER Completed Treatment ...

Page 20: ...uctions table of contents symbols labels limited warranty prev next prev topic next topic Quick Reference Dual Phase Treatment Instructions Select Dose UP DOWN ARROW KEYS ENTER During Treatment Select Polarity UP ARROW KEY ENTER Pre Launch UP ARROW KEY OR ENTER Completed Treatment ...

Page 21: ...s treatments Malignancy The use of Iontophoresis is contraindicated for patients with clinically diagnosed cancer Epilepsy Iontophoresis is contraindicated for patients with Epilepsy Previous Reactions to Electrotherapy The use of Iontophoresis is contraindicated for patients who have had previous reactions to any type of electrotherapy Previous Adverse Reactions Iontophoresis Drugs The use of Ion...

Page 22: ...reatment and consult the prescribing physician 5 Use extreme caution when administering a treatment where sensory nerve damage is present or in any case where there is a loss of normal skin sensation this includes areas desensitized by medication WARNING A transient skin reaction can sometimes occur directly under an electrode This reaction is identified by a uniform red pattern and will usually d...

Page 23: ...soap is recommended Do not immerse the device in water or use cleaning sprays Do not attempt to sterilize the device using any type of sterilization equipment including autoclaves 5 Damaged electrode cables Avoid stretching cables to full length bending cables sharply or wrapping cables tightly Undue stress on cables can damage connections If electrode cables are damaged they should be replaced im...

Page 24: ...Treatment Deliverable Dose 1 160 mA min Number of 40 mA min Treatments 89 Minimum Treatment Time 15 sec Maximum Treatment Time 100 minutes Timing Accuracy 1 second Dimensions Length 4 5 Width 2 66 Height 1 25 Weight 3 6 oz without battery or electrode cable Display Organic Light Emitting Diodes OLED Environmental Conditions Transport and Storage a Ambient temperature range 10C 50C b Relative humid...

Page 25: ...ent specified below The customer or the user of the Dynatron ibox should assure that it is used in such an environment Emissions test Compliance Electromagnetic Environment Guidance RF emissions CISPR 11 EN55011 Group 1 The Dynatron ibox RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF em...

Page 26: ... power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode Not Applicable Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 Ut 95 dip in Ut for 0 5 cycle 40 Ut 60 dip in Ut for 5 cycles 7...

Page 27: ... RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Conducted RF IEC 61000 4 6 Radiated RF EN55011 Not Applicable 3 V m 80 MHz to 2 5 GHz Not Applicable 3 V m NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 Thes...

Page 28: ... communications equipment Separation distance according to frequency of transmitter W Separation distance according to frequency of transmitter 150 kHz to 80 MHz d 1 17 x P 80 MHz to 800 MHz d 1 17 x P 800 MHz to 2 5 GHz d 2 33 x P 0 01 0 12 0 12 0 23 0 01 0 37 0 37 0 74 1 1 2 1 2 2 3 10 3 7 3 7 7 4 100 12 12 23 For transmitters at a maximum output power listed above the recommended separation dis...

Page 29: ...ccur with older batteries that drain quickly during a treatment Long Period Non Use of Batteries If the Dynatron ibox will not be used for a long period of time more than one month it is recommended that the batteries be removed from the battery compartment and stored for future use Replacing the Battery 1 Remove the battery compartment cover located on the back of the device by sliding the cover ...

Page 30: ...w the device to be properly tracked through the repair process and returned to you as quickly as possible Returning a Device for Repair If it becomes necessary to return the device during the warranty period the following information must be supplied when calling Dynatronics Customer Service to obtain a return Service Order Number 1 User name and address 2 User phone number 3 Serial number of the ...

Page 31: ... and then wipe the device off Solvents caustic solutions and harsh or abrasive cleaners must never be used Cleaning the Battery Contacts Battery contacts may be cleaned with a cotton swap lightly dampened with isopropyl alcohol Never use sandpaper or other abrasive material Cleaning the Electrode Cables Periodically clean the electrode cables with a damp cloth and mild soap Gently wipe the cable d...

Page 32: ...of the following symbols are included in the labeling for this device Definitions accompany each symbol Warning Caution On Off power connection to the mains Read the Instructions Follow Instructions for Use Keep Dry Non ionizing Electromagnetic Radiation Temperature Model Number Start Continue Dynatronics Manufacturing Date Serial Number Type BF patient applied part REF Electrical Ratings Label Se...

Page 33: ...ONSTITUTE THE SOLE AND EXCLUSIVE REMEDY IN THE EVENT OF A BREACH OF WARRANTY REGISTRATION REQUIRED In order for this warranty to be valid the warranty registration page included with the product must be filled out and returned to DYNATRONICS within 30 days of purchase by the original owner A copy of an invoice or receipt may be requested to verify purchase date REPAIRS All repairs must be performe...

Page 34: ...covered by this limited warranty ANY IMPLIED WARRANTIES COVERING THIS PRODUCT INCLUDING BUT NOT LIMITED TO WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE ARE LIMITED IN DURATION TO ONE YEAR FROM THE DATE OF PURCHASE BY THE ORIGINAL OWNER DYNATRONICS SHALL NOT IN ANY CASE BE LIABLE FOR SPECIAL INCIDENTAL CONSEQUENTIAL INDIRECT OR OTHER SIMILAR DAMAGES ARISING FROM BREACH OF WARRA...

Page 35: ...________________________________________________________ City_________________________ State_____________ Zip_________________ Dynatronics Dealer__________________________________________________ Salesperson________________________________________________________ I have read and understand the information contained in the operator s manual for this device I have received inservice training from my...

Reviews: