Dynatron® ibox™ Operating Instructions
Custom Treatments
“Custom Treatments” are treatments in which the user explicitly
sets the polarity/mode, and dose(s) of a treatment. The following
on-going steps describe how to setup these parameters and deliver a
custom treatment.
10.
Define a custom treatment from the Treatment Setup Screen
From the Treatment Setup Screen, a custom treatment may be defined
by pressing the ENTER key of the channel from which the custom
treatment will be administered.
11.
Enter a custom polarity/mode
On the start of the definition of a custom treatment, the following
screens may be seen and modified by the user. In these screens, the
user chooses between a “single phase” treatment of specified polarity
or a “dual phase” treatment with starting “+” polarity followed by
a “-” polarity treatment. The up or down arrow keys will scroll the
user through the three choices. When the desired polarity/mode is
finally chosen, the user presses the channel’s ENTER key to accept this
polarity/mode.
12.
Enter a custom dose(s)
For a single phase treatment, use the up
and down arrow keys to adjust the dose
of the treatment to the desired amount.
The dose can be adjusted from 1 to 160
mA-min. Holding down an up or down
arrow key accelerates the dose value up
or down so that the user does not have to
press a key for each increment of dose.
When the desired dose is finally chosen,
press the ENTER key to accept the dose.
A single phase dose entry display is
shown to the right.
For a dual phase treatment, use the up
and down arrow keys to first adjust the
dose of the highlighted positive treatment
to the desired amount and then press the
ENTER key to adjust the negative dose amount (see Dual Dose Entry
illustration). These doses can be adjusted from 1 to 160 mA-min.
After the negative dose is chosen, press the ENTER key to accept
both doses.
13.
Treatment pre-launch
Upon completion of dose entry and
depression of the ENTER key, the device
will enter “treatment pre-launch”. See
treatment pre-launch display illustration
to the right.
In treatment pre-launch, the device
awaits for the user to press either the up
arrow key to start current delivery to the
Single Phase
Negative Polarity
Dual Phase
Polarity
Single Phase
Positive Polarity
Pre-Launch
Dual Phase
Dose Entry
Single Phase
Dose Entry