background image

Long-Term Care Bed • D500 • 5 Function Ultra Low Bed (Reorder No. 12005)

Long-Term Care Bed • D500 • 5 Function Ultra Low Bed (Reorder No. 12005) 

3

WARNINGS

WARNINGS 

(Continued)

DECLARATION – ELECTROMAGNETIC EMISSIONS

The D500 Long-Term Care Bed is intended for use in the 

electromagnetic environment in which radiated RF disturbanc-

es are controlled. The customer or the user of the D500 

Long-Term Care Bed can help prevent electromagnetic interfer-

ence by maintaining a minimum distance between portable and 

mobile RF communications equipment (transmitters) and the 

D500 Long-Term Care Bed and according to the maximum 

output power of the communications equipment. Please contact 

Dynarex at [email protected] for more information 

and manufacturer technical specifications.

Special considerations should be given to proximity of D500 

Long-Term Care Bed and patients who have a cardiac pace-

maker, implanted defibrillator, or other implanted metallic or 

electronic device, because this may cause electrical interfer-

ence, or death. Contact Dynarex [email protected] 

for manufacturer electromagnetic interference technical 

specifications.

Interference to electronic equipment may occur  

in the vicinity of devices marked with this symbol:

•  To engage the Trendelenburg or Reverse Trendelenburg 

position, raise the bed to its highest position first before 

making the adjustment to position. Ensure there are no 

obstruction around the bed before adjusting the position.

WHEN USING RAILS OR ASSIST BARS:

•  Before operating, ensure that the bed rails/bars are  

assembled and installed correctly as instructed. 

• 

DO NOT 

apply side pressure to the bed rails/bars as it can 

deform or damage the bed and/or bed rails/bars. Serious 

harm or death may result from improper use of the bed 

rails/bars.

• 

DO NOT

 use the bed rails/bar as a push handle for moving 

the bed. This could deform or break the bed rails.

• 

DO NOT

 use the bed rails/bar if proper installation cannot 

be achieved. Contact your dealer or manufacturer.

• When lowering the bed below 12.5" ensure the bed rails/

bars are raised. Lowering the bed below 12.5" with the bed 

rails/bars in a lowered position may damage the bed. 

•  Read all instructions before assembling or using the bed.

• 

DO NOT

 exceed maximum weight limitation of the bed. 

The maximum safe working load (SWL) of the bed is  

400 lb. The suggested patient weight is 320 lb. but  

varies based on the accessories on the bed.   

•  The beds are intended for users who fully understand  

the content of this manual and is not intended for use  

by children. 

• 

DO NOT

 operate your bed until it is fully assembled and 

checked. To avoid injury, 

DO NOT

 attempt to remove the 

bed from the carton without assistance.

• 

DO NOT

 use accessories that are not designed or ap-

proved for use with the bed. Use only Dynarex approved 

parts and accessories. 

DO NOT

 make modifications to the 

bed without authorization from the manufacturer. Doing so 

will void the warranty. Serious harm or death may result 

from the use of improper parts or accessories. 

•  The bed is 

NOT

 a transport device. The wheels on the bed 

are for positioning the bed only.

•  Ensure that the casters of the bed are locked at all times 

to prevent unintended bed movement. The bed should only 

be unlocked when the bed is moved. If bed is left unlocked 

unintended movement may occur which may cause dam-

age to the resident or property damage.

•  Moving the bed while the caster locks are engaged may 

damage the bed or the floor. Ensure all the casters are 

unlocked prior to moving the bed.

•  Never permit more than one person on the bed at one 

time.  

•  Ensure that the individual is properly positioned. The 

body weight must be evenly distributed over the sleeping 

surface. No body part should protrude outside the sleep 

surface. Caution should be used when transferring the 

individual to and from the bed.

• 

ALWAYS

 keep hands and feet away from the bottom of 

the bed and any moving parts to avoid injury. Due to the 

low bed clearance, extreme caution should be used as 

under bed space and/or moving parts can create crush or 

pinch points.

• 

DO NOT

 operate the bed with any items under the bed.

• 

ONLY

 use appropriate mattresses and 

ALWAYS

 adjust 

the mattress retainer system accordingly to ensure the 

mattress is fixed in place. The mattress must fit the frame 

properly to avoid any entrapment issues. The wrong  

size mattress may cause serious harm or death.  

For more information about entrapment, visit  

https://www.fda.gov/media/71460/download.

•  Inspect the casters every six months to check for tightness 

and wear.

• 

DO NOT

, under any circumstances, cut or remove the 

ground prong from any plug. Some devices are equipped 

with three-prong (ground) plugs for protection against 

possible shock hazards. Where a two-prong wall recep-

tacle is encountered, it is the personal responsibility and 

obligation of the customer to contact a qualified electrician 

to have the two-prong receptacle replaced with a properly 

grounded three-prong wall receptacle in accordance with 

the regional and national regulations. If you must use an 

extension cable, use 

ONLY

 a three-wire extension cable 

having the same or higher electrical rating as the device 

being connected.

•  Cables should be routed and secured properly to ensure 

they are not damaged during normal operation. 

DO NOT

 

use if any cable is cut, frayed or loosely connected to 

the device. Cables may be damaged by inappropriate 

handling, e.g. by kinking, shearing or other mechanical 

damages. The cables are only replaceable by authorized 

Dynarex service personnel.

•  When moving the bed ensure the power cord and pen-

dants are secured. Ensure that the casters do not roll over 

the cables.

• 

DO NOT

 attempt to open the pre-sealed actuator or obtain 

local service, for it will 

VOID

 the warranty and might result 

in damage. Consult your dealer or manufacturer for further 

information.

•  This bed is not intended to be used in an oxygen rich  

environment.

• 

DO NOT

 allow any part of the bed to become wet or sub-

merged in water.

• 

DO NOT

 use near fire or explosive gases.

•  The hand pendant cord may be a source of entanglement. 

Patients with decreased mental acuity should not have  

access to the pendant as serious harm or death may  

result. An optional staff control embedded into the foot-

board is available.

•  For every 2 minutes of continuous use of the actuator, the 

bed actuator must not be in use for 18 minutes.

•  The use of the Trendelenburg and Reverse Trendelenburg 

may have adverse medical effects. These functions should 

be used under the supervision of medical personnel.  

•  To reduce the possibility of damage to walls or other prop-

erty, the ensure at least a 18" between the wall or other 

objects and the headboard and footboard before engaging 

the Trendelenburg / Reverse Trendelenburg positions.

Summary of Contents for 12005

Page 1: ...Manufactured for Dynarex Corporation 10 Glenshaw Street Orangeburg NY 10962 USA www dynarex com Made in China Long Term Care Bed D500 5 Function Ultra Low Bed User Manual Reorder No 12005...

Page 2: ...unlocked unintended movement may occur which may cause dam age to the resident or property damage Moving the bed while the caster locks are engaged may damage the bed or the floor Ensure all the cast...

Page 3: ...dant Clip 1 M10 Hex Bolt 16 Transport Bar 2 Headboard 1 OPERATING RANGE Bed Height 2 5 25 6 Back Rest Angle 0 70 Knee Break Angle 0 27 Additional Manual Leg Lift Yes various positions Trendelenburg Re...

Page 4: ...two socket screws on each side 4 Tilt the assembly back on its foot 5 Repeat the same procedure for foot board installation 2 3 ASSEMBLING THE BED DECK 1 With both the head and foot end assemblies st...

Page 5: ...stalling the backup battery 2 Affix the backup battery with nuts and bolts to the left of the control box on the same holding plate 3 Open the control box locking lid by pressing in the sides and lift...

Page 6: ...identified and rectified the control system will require a reset in order to operate again If the fault occurred on the motors for platform profiling lower the corresponding motor to the lowest positi...

Page 7: ...ndant buttons are pressed Adjustment may be at the maximum or minimum position Power cord not connected or damaged Hand pendant or actuator connectors are loose Load is too heavy Control system has no...

Page 8: ...roducts and our assurance to the consumer of years of dependable service The Warranty shall not apply under the following conditions Problems arising from normal wear Problems arising from failure to...

Page 9: ...16 Long Term Care Bed D500 5 Function Ultra Low Bed Reorder No 12005 SERVICE RECORD DATE PERFORMED BY CONDITION REPORT...