XB-10/BIOmaster/VIS-20/ENVmaster
instruction manual
32
CONC concentration regression and parameter setting
(
II
)
If using standard sample regression calculation method
(
CALIBRATION , select STD
)
, it will show this
parameter setting interface.
STD1~STD9: 9 standard sample concentration, concentration must be sorted (in general, sort ascending).
If using regression equation parameters to enter calculation method
(
CALIBRATION, select K-Factory
)
,it
will show this parameter setting interface.
K1
Slope value of regression equation
K0
Intercept value of regression equation
OK
Finish parameter setting, back to measure interface automatically
BACK Back to the prior setting interface