11
Dymax ACCU-CAL™ 50-LED Radiometer User Guide
3-mm Lightguide
- For use when measuring energy at the end of a 3-mm lightguide.
5-mm Lightguide
- For use when measuring energy at the end of a 5-mm lightguide.
8-mm Lightguide
- For use when measuring energy at the end of an 8-mm lightguide.
Figure 9. Lightguide Source Screens (Shown in Peak Intensity Mode)
10.
Press and release the Mode button to select the operating mode. Repeatedly press the Mode button
until the mode you desire appears on the second line of the display. The different operating modes are:
Peak Intensity -
Shows the highest intensity in mW/cm
2
seen by the detector during the
measurement and stores that value in the display.
Intensity -
Shows the intensity in mW/cm
2
at the detector during the measurement. This is a dynamic
readout of intensity.
Dose -
Shows the total energy dose in mJ/cm
2
at the detector during the measurement.
Note: This measurement is time dependent.
11.
Turn on the spot lamp to begin testing.