17
Dymax BlueWave® 75 User Guide
Bulb Replacement
1.
Unplug the unit. Remove the top cover from the
BlueWave 75
by loosening the four cover fasteners
(these are the four extended screws on the sides of the top cover).
2.
Unplug the Bulb Connector and loosen the two knurled screws shown in Figure 16. Pivot the white Bulb
Clamp up and out of the way to allow removal of the Bulb.
Figure 16. Bulb Replacement Step 1-2
3.
Gently tip the back end of the Bulb down to un-seat the top edge and remove the Bulb.
Figure 17. Bulb Replacement Step 3
4.
Using powder free gloves, unpack the new Bulb (PN 40205). Take care not to bend the Flat Electrode in
the center of the Bulb.
5.
Orient the new Bulb so the Side Electrode Wire is positioned at the side of the housing that is open.
Carefully seat the bottom edge (Figure 18). The lower Retaining Spring Wire will exert resistance that you
will need to overcome as you seat the Bulb. With the lower edge seated, tip the top edge into the upper
seat. The Bulb will remain in place if properly installed.
Screw
Screw
Bulb Clamp
Bulb
Connector