Dwyer Instruments GFM2 Series Installation And Operating Instructions Manual Download Page 4

1. UNPACKING THE GFM2 MASS FLOW METER

1.1 - Inspect Package for External Damage

Your GFM2 Mass Flow Meter was carefully packed in a sturdy cardboard carton,
with  anti-static  cushioning  materials  to  withstand  shipping  shock.  Upon  receipt,
inspect the package for possible external damage. In case of external damage to
the package contact the shipping company immediately.

1.2 - Unpack the Mass Flow Meter

Open  the  carton  carefully  from  the  top  and  inspect  for  any  sign  of  concealed
shipping  damage.  In  addition  to  contacting  the  shipping  carrier  please  forward  a
copy of any damage report to your distributor or Dwyer Instruments, Inc directly.
When  unpacking  the  instrument  please  make  sure  that  you  have  all  the  items
indicated on the Packing List. Please report any shortages promptly.

1.3 - Maintenance/Repair

Upon final installation of the Series GFM2, no routine maintenance is required. The
Series GFM2 is not field serviceable and should be returned if repair is needed.
Field repair should not be attempted and may void warranty.

Warranty/Return

Refer to “Terms and Conditions of Sales” in our catalog and on our website. Contact
customer service to receive a Return Goods Authorization number before shipping
the product back for repair. Be sure to include a brief description of the problem
plus any additional application notes.

2. INSTALLATION
2.1 - Primary Gas Connections

Please  note  that  the  GFM2  Mass  Flow  Meter  will  not  operate  with  liquids.  Only
clean  gases  are  allowed  to  be  introduced  into  the  instrument.  If  gases  are
contaminated they must be filtered to prevent the introduction of impediments into
the sensor.

Attitude  limit  of  Mass  Flow  Meter  is  ±15°  from  calibration  position  (standard
calibration is in horizontal position). This means that the gas flow path of the Flow
Meter must be within this limit in order to maintain the original calibration accuracy.
Should there be need for a different orientation of the meter, re-calibration may be
necessary.  It  is  also  preferable  to  install  the  GFM2  transducer  in  a  stable
environment, free of frequent and sudden temperature changes, high moisture, and
drafts.

Prior to connecting gas lines inspect all parts of the piping system including ferrules
and fittings for dust or other contaminants. Be sure to observe the direction of gas
flow as indicated by the arrow on the front of the meter when connecting the gas
system to be monitored. Insert tubing into the compression fittings until the ends of
the  properly  sized  tubing  home  flush  against  the  shoulders  of  the  fittings.
Compression  fittings  are  to  be  tightened  according  to  the  manufacturer's
instructions to one and one quarter turns. Avoid over tightening which will seriously
damage the Restrictor Flow Elements (RFE's)!

GFM2  transducers  are  supplied  with  standard  1/4˝  or  3/8˝  inlet  and  outlet
compression  fittings  which  should  not  be  removed  unless  the  meter  is  being
cleaned or calibrated for a new flow range.

Using a Helium Leak Detector or other equivalent method perform a thorough
leak  test  of  the  entire  system.  (All  GFM2’s  are  checked  prior  to  shipment  for
leakage within stated limits. See specifications in this manual.)

2.2 - Electrical Connections

GFM2 is supplied with a 25 pin "D" connector. Pin diagram is presented in figure 
b-1.

2.2.1 - Power Supply Connections

The power supply requirements for GFM2 transduers are: 11 to 26 VDC, (unipolar
power supply)

DC Power (+)               --------------- pin 7 of the 15 pin "D" connector
DC Power (-)                --------------- pin 5 of the 15 pin "D" connector

2.2.2  - Output Signals Connections

GFM2 Mass Flow Meters are equipped with either calibrated 0 to 5 VDC (0 to 10
VDC  optional)  or  calibrated  4  to  20  mA  output  signals  (jumper  selectable).  This
linear output signal represents 0 to 100% of the flow meter's full scale range. 

Flow 0 to 5 VDC or 4 to 20 mA output signal connection:

Plus (+)

------------------- pin 2 of the 15 pin "D" connector

Minus (-)

------------------- pin 1 of the 15 pin "D" connector

To eliminate the possibility of noise interference, use a separate cable entry for the
DC power and signal lines.

2.2.3 - Communication Parameters and Connections

The  digital  interface  operates  via  RS-485  (optional  RS-232  or  Profibus  DP  is
available)  and  provides  access  to  applicable  internal  data  including:  flow,  CPU
temperature, pressure reading, auto zero, totalizer and alarm settings, gas table,
conversion  factors  and  engineering  units  selection,  dynamic  response
compensation and linearization table adjustment.

Communication Settings:

Baud rate:

--------

9600 baud 

Stop bit:

--------

1

Data bits:

--------

8

Parity: 

--------

None

Flow Control:

--------

None

RS-485 Communication Interface Connection:

The  RS-485  converter/adapter  has  to  be  configured  for:  multidrop,  2  wire,  half
duplex  mode.  Settings  for  the  receiver  circuit  usually  should  follow  the  selection
made for the transmitter circuit in order to eliminate echo.

RS-485 T(-) or R(-)             -------- 

pin 8 of the 15 pin "D" connector (TX-)

RS-485 T(+) or R(+)            --------

pin 15 of the 15 pin "D" connector (RX+)

RS-485 GND (if available)    --------

pin 9 of the 15 pin "D" connector (GND)

RS-232 Communication Interface Connection:

Crossover connection has to be established:

RS-232 RX 
(pin 2 on the DB9 connector) -------- pin 8 of the 15 pin "D" connector (TX)
RS-232 TX 
(pin 3 on the DB9 connector) -------- pin 15 of the 15 pin "D" connector (RX)
RS-232 GND 
(pin 5 on the DB9 connector) -------- pin 9 of the 15 pin "D" connector (GND)

GFM2 transducers should not be used for monitoring OXYGEN
gas  unless  specifically  cleaned  and  prepared  for  such

application. For more information, contact your distributor or Dwyer.

CAUTION

Do not apply power voltage above 26 VDC. Doing so will cause
GFM2 damage or faulty operation.

CAUTION

When connecting the load to the output terminals, do not
exceed the rated values shown in the specifications. Failure to

do so might cause damage to this device. Be sure to check that the wiring and
the polarity of the power supply is correct before turning the power ON. Wiring
error may cause damage or faulty operation.

CAUTION

All  4  to  20  mA  current  loop  outputs  are  self-powered  (non-
isolated).  Do  not  connect  an  external  voltage  source  to  the

output signals!

Page 3

For  models  GFM2  models  the  maximum  pressure  in  the  gas
line  should  not  exceed  500  PSIA  (34.47  bars).  Applying

pressure above 500 PSIA (34.47 bars) will seriously damage the flow sensor and
may cause serious injury or death.

WARNING

CAUTION

Summary of Contents for GFM2 Series

Page 1: ...thin 2 of actual flow Output Signal Linear 0 to 5 VDC 3000 min load impedance and 4 to 20 mA 500 max loop resistance Relay Rating 1 A 24 VDC Max Particulate Size 5 microns Temperature Limits 32 to 122...

Page 2: ...ass Flow Meter Installation and Operating Instructions Bulletin F GFM2 DWYER INSTRUMENTS INC Phone 219 879 8000 www dwyer inst com P O BOX 373 MICHIGAN CITY INDIANA 46360 U S A Fax 219 872 9057 e mail...

Page 3: ...7 6 2 Flow Path Cleaning 7 6 2 1 Restrictor Flow Element RFE 7 6 2 2 GFM2 Models 7 7 Calibration Procedures 8 7 1 Flow Calibration 8 7 2 Gas Flow Calibration of GFM2 Mass Flow Meters 8 7 2 1 Connecti...

Page 4: ...thorough leak test of the entire system All GFM2 s are checked prior to shipment for leakage within stated limits See specifications in this manual 2 2 Electrical Connections GFM2 is supplied with a 2...

Page 5: ...nterface or optional Profibus DP interface GFM2 flow meters support various functions including programmable flow totalizer low high or range flow alarm automatic zero adjustment activated via local b...

Page 6: ...peration of the meter above 110 of maximum calibrated flow may increase recovery time 5 3 GFM2 Parameters Settings 5 3 1 Engineering Units Settings The GFM2 Mass Flow Meter is capable of displaying fl...

Page 7: ...nitrogen for up to 36 gases are stored in the GFM2 see APPENDIX II In addition provision is made for a user defined conversion factor Conversion factors may be applied to any of the ten gas calibrati...

Page 8: ...to disassemble the sensor If blockage of the sensor tube is not alleviated by flushing through with cleaning fluids please return meter for servicing 6 2 1 Restrictor Flow Element RFE The Restrictor...

Page 9: ...s the sensor Average reading must be in the range 120 10 counts If it is not perform Auto Zero procedure see section 5 3 10 Zero Calibration 7 2 3 Gas Linearization Table Adjustment Gas flow calibrati...

Page 10: ...RS 232 SOFTWARE INTERFACE COMMANDS 8 1 General The standard GFM2 comes with an RS 485 interface For the optional RS 232 interface the start character and two hexadecimal characters for the address mu...

Page 11: ...command can be used only when absolutely no flow thru the meter and no earlier then 6 minutes after power up It can take several minutes to complete Unit will not respond to other commands when this...

Page 12: ...R read current totalizer volume W warm up delay S setting status D disable sets K 1 I internal K factor U user specified factor S status FS mL sec mL min mL hr L sec L min L hr m3 sec m3 min m3 hr f3...

Page 13: ...w measurement LCD display reading does correspond to the correct flow range but 0 5 Vdc output signal does not change always the same reading or around zero LCD display reading and 0 to 5 Vdc output v...

Page 14: ...nsor temperature is too high MCU temperature is too high overload Fatal Error EEPROM or Auto Zero error Solution Flush clean or disassemble to remove impediments or replace the filter screen see secti...

Page 15: ...rent Gas Table Number 0 9 Current Units of Measure 0 22 Alarm Mode E Enabled D Disabled Low Flow Alarm Setting FS 0 Disabled High Flow Alarm Setting FS 0 Disabled Flow Alarm Action Delay 0 3600 sec 0...

Page 16: ...Flow in PFS 0 6 70 0 FS A D Value from Sensor counts Actual Flow in PFS 0 7 80 0 FS A D Value from Sensor counts Actual Flow in PFS 0 8 90 0 FS A D Value from Sensor counts Actual Flow in PFS 0 9 100...

Page 17: ...n 114 C2Cl2F4 Freon 115 C2ClF5 Freon C318 C4F8 Germane GeH4 Germanium Tetrachloride GeCl4 Helium He 50 L min Helium He 1 50 L min Helium He 2 10 50 L min Index 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16...

Page 18: ...butyl Aluminum C4H9 AL Titanium Tetrachloride TiCl4 Trichloro Ethylene C2HCl3 Trimethylamine CH3 3N Tungsten Hexafluoride WF6 Uranium Hexafluoride UF6 Vinyl Bromide CH2CHBr Vinyl Chloride CH2CHCl Xeno...

Page 19: ...APPENDIX IV Component Diagram Top Component Side Bottom Component Side Page 18...

Page 20: ...s required The Series GFM3 is not field serviceable and should be returned if repair is needed Field repair should not be attempted and may void warranty See 1 3 WARRANTY RETURN Refer to Terms and Con...

Page 21: ...wyer Instruments Inc Printed in U S A 5 13 FR RA 444052 00 DWYER INSTRUMENTS INC Phone 219 879 8000 www dwyer inst com P O BOX 373 MICHIGAN CITY INDIANA 46360 U S A Fax 219 872 9057 e mail info dwyer...

Reviews: